Emerson 3051S Specifications Page 189

  • Download
  • Add to my manuals
  • Print
  • Page
    / 222
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 188
179
Reference Manual
00809-0100-4001, Rev JA
Appendix B: Product Certifications
November 2012
Product certifications
N1 ATEX Non-incendive/Type n and Dust
Certification No.: BAS 00ATEX3105X
Applicable Standards: IEC60079-0:2011, EN60079-15:2010, EN60079-31:2009
Markings: II 3 GD, Ex nA IIC Gc T5 (–40 Ta 70 °C),
Ex ta IIIC T50 °C T
500
60°C Da, IP66
1180
Specific Conditions for Safe Use (X):
1.) The apparatus is not capable of withstanding the 500 V insulation test required by
EN60079-15. This must be taken into account when installing the apparatus.
2.) This device contains a thin wall diaphragm. Installation, maintenance, and use shall
take into account the environmental conditions to which the diaphragm will be
subjected. The manufacturer’s instructions for installation and maintenance shall be
followed in detail to assure safety during its expected lifetime. In case of repair, contact
the manufacturer for more information on the dimensions of the flameproof joints.
Japanese certifications
E4 TIIS Flame-Proof
Markings: Ex d IIC T6
I4 TIIS Intrinsic Safety
Certification No.: TC16406
Markings: Ex ia IIC T4
IECEx certifications
E7 IECEx Flame-proof
Certification No.: IECEx KEM 09.0034X
Applicable Standards: IEC60079-0:2011, IEC60079-1:2007-04, IEC60079-26:2006,
Markings: Ex d IIC T5...T6 Ga/Gb, T5 (-50 °C Ta 80 °C)/T6 (-50 °C Ta 65 °C)
Conditions of Certification (X):
1.) This device contains a thin wall diaphragm. Installation, maintenance, and use shall
take into account the environmental conditions to which the diaphragm will be
subjected. The manufacturer's instructions for installation and maintenance shall be
followed in detail to assure safety during its expected lifetime.
Certificate Description
TC15850 3051C/D/1 4–20 mA HART — no meter
TC15851 3051C/D/1 4–20 mA HART — with meter
TC15854 3051T/G/1 4–20 mA HART, SST, Silicon — no meter
TC15855 3051T/G/1 4–20 mA HART, Alloy C-276, Silicon — no meter
TC15856 3051T/G/1 4–20 mA HART, SST, Silicon — with meter
TC15857 3051T/G/1 4–20 mA HART, Alloy C-276, Silicon — with meter
Process
Temp
Ambient
Temp
Temp Class
-50 to 65 -50 to 65 T6
-50 to 80 -50 to 80 T5