Emerson 3051S Specifications Page 102

  • Download
  • Add to my manuals
  • Print
  • Page
    / 188
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 101
12
Reference Manual
00809-0100-4803, Rev DA
Section 4: Operation and Maintenance
April 2013
Operation and Maintenance
Data flow can be summarized in four major steps:
1. A change in a process variable (DP, P, and/or T) corresponds to a change in the sensor
output (Sensor Signal).
2. The sensor signal is converted to a digital format that is understood by the
microprocessor (Analog-to-Digital Signal Conversion).
3. Corrections and flow calculations are performed in the microprocessor to obtain a
digital representation of the process output variables.
4. The Digital Primary Variable (PV) is converted to an analog value (Digital-to-Analog
Signal Conversion).
4.3.2 Sensor trim overview
Trim the sensors using either sensor or zero trim functions. Trim functions vary in complexity
and are application-dependent. Both trim functions alter the transmitter’s interpretation of the
input signal.
Zero trim
Zero trim is a single-point offset adjustment. It is useful for compensating for mounting position
effects and is most effective when performed with the transmitter installed in its final mounting
position. Since this correction maintains the slope of the characterization curve, it should not be
used in place of a sensor trim over the full sensor range.
When performing a zero trim with a manifold, refer to “Rosemount 305 and 304 Manifolds” on
page 19.
Note
The transmitter must be within five percent or less of the maximum span of true zero
(zero-based) in order to calibrate with zero trim function.
The transmitter will not allow the user to perform a zero trim on an absolute static pressure
sensor. To correct mounting position effects on the absolute static pressure sensor, perform a
lower sensor trim. The lower sensor trim function provides an offset correction similar to the
zero trim function, but it does not require zero-based input.
Upper and lower sensor trim
Sensor trim is a two-point sensor calibration where two end-point pressures are applied, and all
output is linearized between them. Always adjust the lower sensor trim value first to establish
the correct offset. Adjustment of the upper sensor trim value provides a slope correction to the
characterization curve based on the lower sensor trim value. The trim values allow the user to
optimize performance over a specified measuring range at the calibration temperature.
Page view 101
1 2 ... 97 98 99 100 101 102 103 104 105 106 107 ... 187 188

Comments to this Manuals

No comments