www.rosemount.comReference Manual 00809-0100-4801, Rev FAOctober 2010Rosemount 3051S SeriesScalable Pressure, Flow, and Level Solutionwith HART® Proto
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S SeriesTOC-6
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series6-2CERTIFICATION The 3051S safety certified pressure transmitter is certified
Reference Manual 00809-0100-4801, Rev FAOctober 20106-3Rosemount 3051S SeriesCOMMISSIONING To commission the 3051S Safety Certified Transmitter, use t
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series6-4Switches installed1. If using a communicator, use the following fast key s
Reference Manual 00809-0100-4801, Rev FAOctober 20106-5Rosemount 3051S SeriesOPERATION AND MAINTENANCEProof Test The following proof tests are recomm
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series6-6Inspection Visual InspectionNot requiredSpecial ToolsNot requiredProduct R
Reference Manual 00809-0100-4801, Rev FAOctober 20107-1Rosemount 3051S SeriesSection 7 Advanced HART Diagnostic SuiteOverview . . . . . . . . . . .
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series7-21. Statistical Process Monitoring (SPM) – SPM technology detects changes i
Reference Manual 00809-0100-4801, Rev FAOctober 20107-3Rosemount 3051S SeriesUSER INTERFACE The 3051S with Advanced HART Diagnostic Suite can be used
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series7-4STATISTICAL PROCESS MONITORINGIntroduction Statistical Process Monitoring
Reference Manual 00809-0100-4801, Rev FAOctober 20107-5Rosemount 3051S SeriesFigure 7-3. CV is the ratio of Standard Deviation to MeanCV is stable if
Reference Manual 00809-0100-4801, Rev FAOctober 20101-1Rosemount 3051S SeriesSection 1 IntroductionUsing This Manual . . . . . . . . . . . . . . . .
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series7-6Figure 7-4. Statistical Process Monitoring diagnostic resident in transmit
Reference Manual 00809-0100-4801, Rev FAOctober 20107-7Rosemount 3051S SeriesFigure 7-5. Simplified SPM flowchartFurther detail of the operation of th
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series7-8In the “Monitoring” mode, statistical values of mean, standard deviation,
Reference Manual 00809-0100-4801, Rev FAOctober 20107-9Rosemount 3051S SeriesFigure 7-6. Selection of statistical values as secondary variablesSPM Con
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series7-10Figure 7-8. Statistical Process Monitoring main screenThe SPM Status scre
Reference Manual 00809-0100-4801, Rev FAOctober 20107-11Rosemount 3051S SeriesFigure 7-9. Baseline Configuration screenLearn/Monitor Period: This is t
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series7-12Power Interruption ActionThis is used to direct what the diagnostic shoul
Reference Manual 00809-0100-4801, Rev FAOctober 20107-13Rosemount 3051S SeriesThe Detection Configuration screen (Figure 7-11 and Figure 7-12) allows
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series7-14Figure 7-12. Detection Configuration screen for Coefficient of VariationS
Reference Manual 00809-0100-4801, Rev FAOctober 20107-15Rosemount 3051S SeriesThreshold ValueIf sensitivity is Custom, this field will display the cus
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series1-2MODELS COVERED The following 3051S pressure transmitters and the Rosemount
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series7-16Low Detection MessageCustomizable message field related to standard devia
Reference Manual 00809-0100-4801, Rev FAOctober 20107-17Rosemount 3051S SeriesResetIf SPM trip action is set to “Alert Latched”, clicking on Reset wil
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series7-18The LCD will also indicate the diagnostic condition. The “Time Since Dete
Reference Manual 00809-0100-4801, Rev FAOctober 20107-19Rosemount 3051S SeriesNOTERosemount cannot absolutely warrant or guarantee that Statistical Pr
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series7-20POWER ADVISORYIntroduction The Power Advisory diagnostic provides a means
Reference Manual 00809-0100-4801, Rev FAOctober 20107-21Rosemount 3051S SeriesOverview The transmitter is shipped with Power Advisory off as default a
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series7-22Figure 7-18. Manual configuration of Power Advisory main screenThe Power
Reference Manual 00809-0100-4801, Rev FAOctober 20107-23Rosemount 3051S SeriesResistanceThis value is the calculated resistance of the electrical loop
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series7-24TroubleshootingDIAGNOSTIC LOGOverview The Diagnostic Log provides a histo
Reference Manual 00809-0100-4801, Rev FAOctober 20107-25Rosemount 3051S SeriesFigure 7-20. Diagnostic LogFigure 7-20 shows the Diagnostic Log screen w
Reference Manual 00809-0100-4801, Rev FAOctober 20101-3Rosemount 3051S SeriesSERVICE SUPPORT To expedite the return process outside of the United Stat
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series7-26NOTE:It is recommended that transmitters showing “Failed” status should b
Reference Manual 00809-0100-4801, Rev FAOctober 20107-27Rosemount 3051S SeriesFigure 7-21. Pressure Variable Logging screenTime Outside Sensor Limits
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series7-28Temperature Variable LogMinimum, Maximum TemperatureThe meter indicates t
Reference Manual 00809-0100-4801, Rev FAOctober 20107-29Rosemount 3051S SeriesReset All Temperature EventsSelecting this will reset all values on this
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series7-30Total Event Time (High / Low)These fields show the total time the transmi
Reference Manual 00809-0100-4801, Rev FAOctober 20107-31Rosemount 3051S SeriesLatched alert action is recommended if 3rd party alert monitor software
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series7-32Figure 7-25. Service Alert screenTime RemainingAmount of time remaining b
Reference Manual 00809-0100-4801, Rev FAOctober 20107-33Rosemount 3051S SeriesFigure 7-26. Device Diagnostics screenmA Output Diagnostic The mA Output
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series7-34Installation and CommissioningBelow are the four major steps to commissio
Reference Manual 00809-0100-4801, Rev FAOctober 20107-35Rosemount 3051S Series1. Check the 3051S variable mapping and remap as necessary to assign the
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series1-4
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series7-36SAFETY INSTRUMENTED SYSTEMS (SIS) CERTIFICATIONThe safety-critical output
Reference Manual 00809-0100-4801, Rev FAOctober 20107-37Rosemount 3051S SeriesFor more information on the 375 Field Communicator see document 00809-01
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series7-38Configuring Alarm and Saturation Levels1. If using a Field Communicator,
Reference Manual 00809-0100-4801, Rev FAOctober 20107-39Rosemount 3051S SeriesComprehensive Proof TestThe comprehensive proof test consists of perform
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series7-40Failure ValuesProduct Life50 years - based on worst case component wear-o
Reference Manual 00809-0100-4801, Rev FAOctober 20107-41Rosemount 3051S SeriesFIELD COMMUNICATOR MENU TREESFigure 7-31. Overview Menu TreeHome1 Overvi
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series7-42Figure 7-32. Configure (Guided Setup and Manual Setup) Menu TreeHome1 Ove
Reference Manual 00809-0100-4801, Rev FAOctober 20107-43Rosemount 3051S SeriesFigure 7-33. Configure (Alert Setup) Menu TreeHome1 Overview2 Configure3
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series7-44Figure 7-34. Service Tools Menu TreeHome1 Overview2 Configure3 Service To
Reference Manual 00809-0100-4801, Rev FAOctober 2010A-1Rosemount 3051S SeriesAppendix A Specifications and Reference DataPerformance Specifications
Reference Manual 00809-0100-4801, Rev FAOctober 20102-1Rosemount 3051S SeriesSection 2 InstallationOverview . . . . . . . . . . . . . . . . . . . . .
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S SeriesA-2Transmitter Total PerformanceTotal performance is based on combined errors
Reference Manual 00809-0100-4801, Rev FAOctober 2010A-3Rosemount 3051S SeriesWarranty(1)Dynamic PerformanceModels Ultra and Ultra for Flow ClassicAll
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S SeriesA-4Ambient Temperature EffectTransmitter with Coplanar Sensor Module (Single
Reference Manual 00809-0100-4801, Rev FAOctober 2010A-5Rosemount 3051S SeriesMounting Position EffectsVibration Effect Less than ±0.1% of URL when tes
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S SeriesA-6FUNCTIONAL SPECIFICATIONSRange and Sensor LimitsTransmitter with Coplanar
Reference Manual 00809-0100-4801, Rev FAOctober 2010A-7Rosemount 3051S SeriesMinimum Span LimitsTransmitter with Coplanar Sensor Module (Single Variab
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S SeriesA-8Load Limitations Maximum loop resistance is determined by the voltage leve
Reference Manual 00809-0100-4801, Rev FAOctober 2010A-9Rosemount 3051S SeriesOverpressure Limits Transmitters withstand the following limits without d
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S SeriesA-10Temperature Limits AmbientStorageProcess Temperature LimitsAt atmospheric
Reference Manual 00809-0100-4801, Rev FAOctober 2010A-11Rosemount 3051S SeriesHumidity Limits 0–100% relative humidityTurn-On Time When power is appli
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series2-2CONSIDERATIONSGeneral Measurement performance depends upon proper installa
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S SeriesA-12PHYSICAL SPECIFICATIONSSafety-Certified Transmitter Failure ValuesSafety
Reference Manual 00809-0100-4801, Rev FAOctober 2010A-13Rosemount 3051S Series3051S_L Mounting FlangeZinc-cobalt plated CS or 316 SST3051S_L Seal Exte
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S SeriesA-14Shipping WeightsSensor Module WeightsTransmitter Option WeightsCoplanar S
Reference Manual 00809-0100-4801, Rev FAOctober 2010A-15Rosemount 3051S Series3051S_L Weights Without SuperModule Platform, Housing, or Transmitter Op
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S SeriesA-16DIMENSIONAL DRAWINGSFigure A-1. Transmitter with Coplanar Sensor Module a
Reference Manual 00809-0100-4801, Rev FAOctober 2010A-17Rosemount 3051S SeriesFigure A-2. Transmitter with Coplanar Sensor Module and Traditional Flan
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S SeriesA-18Figure A-3. Transmitter with In-Line Sensor ModulePlantWeb Housing Juncti
Reference Manual 00809-0100-4801, Rev FAOctober 2010A-19Rosemount 3051S SeriesFigure A-4. Coplanar Mounting Configurations (B4 Bracket)Pipe Mount Pane
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S SeriesA-20Figure A-5. Traditional Mounting ConfigurationsFigure A-6. In-Line Mounti
Reference Manual 00809-0100-4801, Rev FAOctober 2010A-21Rosemount 3051S SeriesFigure A-7. Remote Display Mounting Configurations (B4 Bracket)(1) Tole
Reference Manual 00809-0100-4801, Rev FAOctober 20102-3Rosemount 3051S SeriesSide MountingWhen the transmitter is mounted on its side, position the Co
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S SeriesA-22Figure A-8. Rosemount 3051S_L Liquid Level TransmitterSide View Front Vie
Reference Manual 00809-0100-4801, Rev FAOctober 2010A-23Rosemount 3051S SeriesORDERING INFORMATIONTable A-1. Rosemount 3051S Scalable Coplanar Pressu
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S SeriesA-24E11 Coplanar flange1/4–18 NPT CS 316 SST ★E12 Coplanar flange1/4–18 NPT S
Reference Manual 00809-0100-4801, Rev FAOctober 2010A-25Rosemount 3051S SeriesExpanded1C PlantWeb housing Aluminum G1/21L PlantWeb housing SST G1/22C
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S SeriesA-26Hardware AdjustmentsStandard StandardD1(13)(17)(18)Hardware adjustments (
Reference Manual 00809-0100-4801, Rev FAOctober 2010A-27Rosemount 3051S SeriesSensor Fill FluidStandard StandardL1(24)Inert sensor fill fluid ★O-ring
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S SeriesA-28Conduit Electrical Connector Standard StandardGE(34)M12, 4-pin, Male Conn
Reference Manual 00809-0100-4801, Rev FAOctober 2010A-29Rosemount 3051S SeriesTable A-2. Rosemount 3051S Scalable In-Line Pressure Transmitter Orderi
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S SeriesA-30Expanded1C PlantWeb housing Aluminum G1/21L PlantWeb housing SST G1/22C J
Reference Manual 00809-0100-4801, Rev FAOctober 2010A-31Rosemount 3051S SeriesProduct Certifications(15) Standard StandardE1 ATEX Flameproof ★I1 ATEX
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series2-4Figure 2-1. HART Installation FlowchartSTART HEREBench Calibration?Field I
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S SeriesA-32Quality Certification for SafetyStandard StandardQS(8)(11)Prior-use certi
Reference Manual 00809-0100-4801, Rev FAOctober 2010A-33Rosemount 3051S Series Table A-3. Rosemount 3051S Liquid Level Transmitter Ordering Informati
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S SeriesA-34Seal System TypeStandard Standard1 Direct-mount seal system ★High Pressu
Reference Manual 00809-0100-4801, Rev FAOctober 2010A-35Rosemount 3051S SeriesLower Housing Material for FF, Extension Length for EF (High Side)(6)Flu
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S SeriesA-36PlantWeb Control FunctionalityStandard StandardA01(11)FOUNDATION fieldbus
Reference Manual 00809-0100-4801, Rev FAOctober 2010A-37Rosemount 3051S SeriesI5 FM Intrinsically Safe, Division 2 ★I6 CSA Intrinsically Safe ★I7 IECE
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S SeriesA-38Material Traceability CertificationStandard StandardQ8 Material traceabil
Reference Manual 00809-0100-4801, Rev FAOctober 2010A-39Rosemount 3051S SeriesTable A-4. Housing Kit for Rosemount 3051S Series Ordering Information★
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S SeriesA-40K5 FM Explosion-proof, Dust Ignition-proof, Intrinsically Safe, Division
Reference Manual 00809-0100-4801, Rev FAOctober 2010A-41Rosemount 3051S SeriesEXPLODED VIEW DIAGRAMThe following drawing shows the name and location f
Reference Manual 00809-0100-4801, Rev FAOctober 20102-5Rosemount 3051S SeriesINSTALLATION PROCEDURESFor dimensional drawing information refer to Appen
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S SeriesA-42SPARE PARTSSee Rosemount 3051S_C, 3051S_T & 3051S_L ordering tables i
Reference Manual 00809-0100-4801, Rev FAOctober 2010A-43Rosemount 3051S SeriesRemote Meter Terminal BlocksPlantWeb Housing 7-Position Remote Communica
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S SeriesA-44Cast Alloy 400 Adapters 03031-1300-0104Ni Plated CS Adapters 03031-1300-0
Reference Manual 00809-0100-4801, Rev FAOctober 2010A-45Rosemount 3051S SeriesBolt KitsCOPLANAR FLANGEFlange Bolt Kit {44 mm (1.75 in.)}Carbon Steel (
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S SeriesA-46
Reference Manual 00809-0100-4801, Rev FAOctober 2010B-1Rosemount 3051S SeriesAppendix B Product CertificationsApproved Manufacturing Locations . . .
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S SeriesB-2HAZARDOUS LOCATIONS CERTIFICATIONSNorth American CertificationsFM Approval
Reference Manual 00809-0100-4801, Rev FAOctober 2010B-3Rosemount 3051S SeriesSpecial conditions for safe use (x)1. The apparatus, excluding the Types
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S SeriesB-4Japanese Certifications E4 TIIS FlameproofEx d IIC T6China (NEPSI) Certifi
Reference Manual 00809-0100-4801, Rev FAOctober 2010B-5Rosemount 3051S Series11. During installation, use, and maintenance of the pressure transmitter
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series2-6Figure 2-2. Cover Jam ScrewMount the Transmitter Mounting BracketsFacilita
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S SeriesB-68. During installation, use, and maintenance of the pressure transmitter,
Reference Manual 00809-0100-4801, Rev FAOctober 2010B-7Rosemount 3051S SeriesINMETRO Certifications I2 Brazilian Approval (INMETRO Approval) - Intrins
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S SeriesB-8IECEx DustCertificate No. IECExBAS09.0014XEx tD A20 T105°C (-20 °C ≤ Tamb
Reference Manual 00809-0100-4801, Rev FAOctober 2010B-9Rosemount 3051S SeriesN7 IECEx Type nCertificate No.: IECExBAS04.0018XEx nC IIC T4 (Ta = -40 °C
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S SeriesB-10INSTALLATION DRAWINGSFactory Mutual (FM)
Reference Manual 00809-0100-4801, Rev FAOctober 2010B-11Rosemount 3051S Series
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S SeriesB-12
Reference Manual 00809-0100-4801, Rev FAOctober 2010B-13Rosemount 3051S Series
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S SeriesB-14
Reference Manual 00809-0100-4801, Rev FAOctober 2010B-15Rosemount 3051S Series
Reference Manual 00809-0100-4801, Rev FAOctober 20102-7Rosemount 3051S SeriesBolt InstallationOnly use bolts supplied with the Rosemount 3051S or sold
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S SeriesB-16
Reference Manual 00809-0100-4801, Rev FAOctober 2010B-17Rosemount 3051S Series
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S SeriesB-18
Reference Manual 00809-0100-4801, Rev FAOctober 2010B-19Rosemount 3051S Series
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S SeriesB-20
Reference Manual 00809-0100-4801, Rev FAOctober 2010B-21Rosemount 3051S Series
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S SeriesB-22
Reference Manual 00809-0100-4801, Rev FAOctober 2010B-23Rosemount 3051S Series
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S SeriesB-24
Reference Manual 00809-0100-4801, Rev FAOctober 2010B-25Rosemount 3051S Series
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series2-8Transmitter withFlange Adaptersand Flange/ Adapter BoltsDescription Qty Si
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S SeriesB-26Canadian Standards Association (CSA)
Reference Manual 00809-0100-4801, Rev FAOctober 2010B-27Rosemount 3051S Series
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S SeriesB-28
Reference Manual 00809-0100-4801, Rev FAOctober 2010B-29Rosemount 3051S Series
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S SeriesB-30
Reference Manual 00809-0100-4801, Rev FAOctober 2010B-31Rosemount 3051S Series
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S SeriesB-32
Reference Manual 00809-0100-4801, Rev FAOctober 2010B-33Rosemount 3051S Series
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S SeriesB-34
Reference Manual 00809-0100-4801, Rev FAOctober 2010B-35Rosemount 3051S Series
Reference Manual 00809-0100-4801, Rev FAOctober 20102-9Rosemount 3051S SeriesImpulse PipingThe piping between the process and the transmitter must acc
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S SeriesB-36
Reference Manual 00809-0100-4801, Rev FAOctober 2010B-37Rosemount 3051S Series
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S SeriesB-38
Reference Manual 00809-0100-4801, Rev FAOctober 2010B-39Rosemount 3051S SeriesKEMA
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S SeriesB-40
Reference Manual 00809-0100-4801, Rev FAOctober 2010B-41Rosemount 3051S Series
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S SeriesB-42
Reference Manual 00809-0100-4801, Rev EAOctober 2010Index-1Rosemount 3051S SeriesAAddressChanging . . . . . . . . . . . 3-31Alarm . . . . . . . . .
Reference Manual00809-0100-4801, Rev EAOctober 2010Rosemount 3051S SeriesIndex-2MMaintenance . . . . . . . . . . . . . 4-1Manifold installations . .
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series2-10Suite includes: Statistical Process Monitoring (SPM), Power Advisory, Sta
Reference Manual 00809-0100-4801, Rev FAOctober 2010Rosemount 3051S SeriesStandard Terms and Conditions of Sale can be found at www.rosemount.com/term
Reference Manual 00809-0100-4801, Rev FAOctober 20102-11Rosemount 3051S SeriesFigure 2-3. Coplanar Installation ExamplesFigure 2-4. In-Line Installati
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series2-12Figure 2-5. O-Rings.Whenever you remove flanges or adapters, visually ins
Reference Manual 00809-0100-4801, Rev FAOctober 20102-13Rosemount 3051S Series1. Loosen the housing rotation set screw.2. Turn the housing up to 180°
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series2-14Figure 2-7. Switch and jumper configuration (option D1)Field Communicator
Reference Manual 00809-0100-4801, Rev FAOctober 20102-15Rosemount 3051S SeriesField Communicator1. Enter the fast key sequence “Local Keys Control” to
Reference Manual 00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Serieswww.rosemount.comRosemount 3051S SeriesScalable Pressure, Flow, and Level So
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series2-16Connect Wiring and Power UpUse twisted pairs to yield best results. To en
Reference Manual 00809-0100-4801, Rev FAOctober 20102-17Rosemount 3051S SeriesSignal Wiring GroundingDo not run signal wiring in conduit or open trays
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series2-18Electrical ConsiderationsProper electrical installation is necessary to p
Reference Manual 00809-0100-4801, Rev FAOctober 20102-19Rosemount 3051S SeriesFigure 2-10. Remote Mount Display wiring diagramNOTEWire colors provided
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series2-20Quick Connect Wiring As standard, the 3051S Quick Connect arrives properl
Reference Manual 00809-0100-4801, Rev FAOctober 20102-21Rosemount 3051S SeriesConduit Electrical Connector Wiring (Option GE or GM)For 3051S transmitt
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series2-22Use the following procedure and Figure 2-13 to install the LCD display:1.
Reference Manual 00809-0100-4801, Rev FAOctober 20102-23Rosemount 3051S SeriesRosemount 305 Integral Manifold Installation ProcedureTo install a 305 I
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series2-247. Subtract the number of threads showing (after tightening) from the tot
Reference Manual 00809-0100-4801, Rev FAOctober 20102-25Rosemount 3051S SeriesManifold Operation Improper installation or operation of manifolds may r
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series2-26Drain/VentValveDrain/Vent ValveEqualize(closed)ProcessIsolate(closed)Isol
Reference Manual 00809-0100-4801, Rev FAOctober 20102-27Rosemount 3051S SeriesFive-valve Natural Gas configurations shown:NOTEDo not open the low side
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series2-283. Open the equalize valve on the low pressure (downstream) side of the t
Reference Manual 00809-0100-4801, Rev FAOctober 20103-1Rosemount 3051S SeriesSection 3 ConfigurationOverview . . . . . . . . . . . . . . . . . . . .
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series3-2COMMISSIONING ON THE BENCH WITH HARTCommissioning consists of testing the
Reference Manual 00809-0100-4801, Rev FAOctober 20103-3Rosemount 3051S SeriesWiring Diagrams Bench Hook-upConnect the bench equipment as shown in Fig
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series3-4REVIEW CONFIGURATION DATANOTEInformation and procedures in this section th
Reference Manual 00809-0100-4801, Rev FAOctober 20103-5Rosemount 3051S SeriesFIELD COMMUNICATOR (Version 3.3)Field Communicator User InterfaceFigure 3
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series3-6Traditional Interface Menu Tree1. Universal Rev.2. Field Device Rev.3. Sof
Reference Manual 00809-0100-4801, Rev FAOctober 20103-7Rosemount 3051S SeriesDevice Dashboard Menu TreeFigure 3-5. 3051S Device Dashboard - OverviewHo
Reference Manual 00809-0100-4801, Rev FAOctober 2010TOC-1Rosemount 3051S SeriesSECTION 1IntroductionUsing This Manual . . . . . . . . . . . . . . . .
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series3-8Figure 3-6. 3051S Device Dashboard - ConfigureHome1 Overview2 Configure3 S
Reference Manual 00809-0100-4801, Rev FAOctober 20103-9Rosemount 3051S SeriesFigure 3-7. 3051S Device Dashboard - Service ToolsHome1 Overview2 Configu
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series3-10Traditional Fast Key SequenceThe following menu indicates fast key sequen
Reference Manual 00809-0100-4801, Rev FAOctober 20103-11Rosemount 3051S SeriesDevice Dashboard Fast Key SequenceThe following menu indicates fast key
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series3-12CHECK OUTPUT Before performing other transmitter on-line operations, revi
Reference Manual 00809-0100-4801, Rev FAOctober 20103-13Rosemount 3051S SeriesBASIC SETUPSet Process Variable UnitsThe PV Unit command sets the proces
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series3-14Figure 3-8. Square Root Output Transition PointNOTEFor a flow turndown of
Reference Manual 00809-0100-4801, Rev FAOctober 20103-15Rosemount 3051S SeriesNOTEIf the transmitter security jumper/switch is ON, adjustments to the
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series3-16Rerange with a Pressure Input Source and the Local Zero and Span buttons
Reference Manual 00809-0100-4801, Rev FAOctober 20103-17Rosemount 3051S SeriesField Communicator v3.3Enter the fast key sequence “Damping.”AMS v7.0Rig
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S SeriesTOC-2SECTION 3ConfigurationOverview . . . . . . . . . . . . . . . . . . . . .
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series3-18DETAILED SETUPFailure Mode Alarm and Saturation3051S transmitters automat
Reference Manual 00809-0100-4801, Rev FAOctober 20103-19Rosemount 3051S SeriesAlarm and Saturation Level ConfigurationTo configure alarm and saturatio
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series3-20Alarm and Saturation Values for Multidrop ModeTransmitters set to multidr
Reference Manual 00809-0100-4801, Rev FAOctober 20103-21Rosemount 3051S SeriesAMS v7.0Right click on the device and select “Configure” from the menu.1
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series3-22Field Communicator v3.3To configure the scaled variable with a Field Comm
Reference Manual 00809-0100-4801, Rev FAOctober 20103-23Rosemount 3051S SeriesAMS v7.0 Right click on the device and select “Device Configuration” the
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series3-24DP Flow ExampleA differential transmitter is used in conjunction with an
Reference Manual 00809-0100-4801, Rev FAOctober 20103-25Rosemount 3051S SeriesAMS v7.0 Right click on the device and select “Configure”.1. In “Basic S
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series3-26Loop Test The Loop Test command verifies the output of the transmitter, t
Reference Manual 00809-0100-4801, Rev FAOctober 20103-27Rosemount 3051S Series5. Check the reference meter installed in the test loop to verify that i
Reference Manual 00809-0100-4801, Rev FAOctober 2010TOC-3Rosemount 3051S SeriesSECTION 4Operation and MaintenanceOverview. . . . . . . . . . . . . . .
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series3-285. Locate the saved transmitter configuration file.a. Select “Offline” b.
Reference Manual 00809-0100-4801, Rev FAOctober 20103-29Rosemount 3051S SeriesAMS v7.0 Applying a User ConfigurationAny amount of user configurations
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series3-30AMS v7.0Right click on the device and select “Configure” from the menu.1.
Reference Manual 00809-0100-4801, Rev FAOctober 20103-31Rosemount 3051S SeriesThe 3051S is set to address zero (0) at the factory, which allows operat
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series3-32
Reference Manual 00809-0100-4801, Rev FAOctober 20104-1Rosemount 3051S SeriesSection 4 Operation and MaintenanceOverview . . . . . . . . . . . . . .
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series4-2Table 4-1. Recommended Calibration TasksNOTE:A Field Communicator is requ
Reference Manual 00809-0100-4801, Rev FAOctober 20104-3Rosemount 3051S SeriesCalibrate the sensor• Sensor Trim (page 4-6)• Zero Trim (page 4-6)Calibra
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series4-4Determining Calibration FrequencyCalibration frequency can vary greatly de
Reference Manual 00809-0100-4801, Rev FAOctober 20104-5Rosemount 3051S SeriesChoosing a Trim ProcedureTo decide which trim procedure to use, you must
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S SeriesTOC-4SECTION 7Advanced HART Diagnostic SuiteOverview . . . . . . . . . . . .
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series4-6Zero Trim NOTEThe transmitter must be within three percent of true zero (z
Reference Manual 00809-0100-4801, Rev FAOctober 20104-7Rosemount 3051S SeriesNOTESelect pressure input values so that lower and upper values are equal
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series4-8Digital-to-Analog Trim Field CommunicatorTo perform a digital-to-analog tr
Reference Manual 00809-0100-4801, Rev FAOctober 20104-9Rosemount 3051S SeriesDigital-to-Analog Trim Using Other ScaleThe Scaled D/A Trim command match
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series4-10Recall Factory Trim—Analog OutputThe Recall Factory Trim—Analog Output co
Reference Manual 00809-0100-4801, Rev FAOctober 20104-11Rosemount 3051S SeriesThe following specifications show the static pressure effect for the 305
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series4-12In this example: To calculate the low trim (LT) value: To calculate the h
Reference Manual 00809-0100-4801, Rev FAOctober 20104-13Rosemount 3051S SeriesDiagnostic Messages In addition to output, the LCD displays abbreviated
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series4-14XMRT INFOA non-volatile memory fault has been detected in the transmitter
Reference Manual 00809-0100-4801, Rev FAOctober 20104-15Rosemount 3051S SeriesFIELD UPGRADESLabeling Each housing and each SuperModule is labeled indi
Reference Manual 00809-0100-4801, Rev FAOctober 2010TOC-5Rosemount 3051S SeriesAPPENDIX ASpecifications and Reference DataPerformance Specifications .
Reference Manual 00809-0100-4801, Rev FAOctober 20104-16Rosemount 3051S Series
Reference Manual 00809-0100-4801, Rev FAOctober 20105-1Rosemount 3051S SeriesSection 5 TroubleshootingOverview . . . . . . . . . . . . . . . . . . .
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series5-2Table 5-1. Rosemount 3051Stroubleshooting tableSymptom Corrective Actions
Reference Manual 00809-0100-4801, Rev FAOctober 20105-3Rosemount 3051S SeriesDISASSEMBLY PROCEDURESDo not remove the instrument cover in explosive atm
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series5-4Remove Interface AssemblyThe Standard Interface Assembly, Adjustment Inter
Reference Manual 00809-0100-4801, Rev FAOctober 20105-5Rosemount 3051S SeriesREASSEMBLY PROCEDURESIMPORTANTThe V-Seal must be installed at the bottom
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series5-6Reassemble the Process Flange1. Inspect the SuperModule PTFE O-rings. If t
Reference Manual 00809-0100-4801, Rev FAOctober 20105-7Rosemount 3051S SeriesTable 5-2. Bolt InstallationTorque Values4. If you replaced the PTFE Sup
Reference Manual00809-0100-4801, Rev FAOctober 2010Rosemount 3051S Series5-8
Reference Manual 00809-0100-4801, Rev FAOctober 20106-1Rosemount 3051S SeriesSection 6 Safety Instrumented SystemsSafety Messages . . . . . . . . . .