Emerson Oxymitter 4000 Specifications Page 101

  • Download
  • Add to my manuals
  • Print
  • Page
    / 146
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 100
IB-106-340
6-3
DIAGNOSTIC
ALARMS
TEST
POINTS
HEATER T/C
HEATER
O2 CELL
CALIBRATION
CALIBRATION RECOMMENDED
O2 CELL mV +
O2 CELL mv -
HEATER T/C +
HEATER T/C -
INC INC
DEC DEC
HIGH
GAS
LOW
GAS
CAL
TEST GAS +
PROCESS -
% O2
22220067
Figure 6-2. Membrane Keypad
(c) If using an IMPS 4000, enter a time
interval via the IMPS 4000 keypad that
will initiate an automatic calibration at
a scheduled time interval (in hours). To
set the CalIntvX parameter of the
CHANGE PRESETS display mode, re-
fer to the IMPS 4000 Intelligent Multi-
probe Test Gas Sequencer Instruction
Bulletin for more information.
Once an automatic calibration is initiated, by
any of the methods previously described, the
Oxymitter 4000s CALIBRATION
RECOMMENDED alarm signals an IMPS
4000 or SPS 4000 to initiate a calibration.
The sequencer sends an in cal signal to the
control room so that any automatic control
loops can be placed in manual. Then, the se-
quencer begins to sequence the calibration
gases.
2. Semi-Automatic Calibration. Semi-
automatic calibrations only require operator
initiation. However, the calibration gases
must be permanently piped to the Oxymitter
4000, an SPS 4000 or IMPS 4000 must be
installed to sequence the gases, and the
Oxymitter 4000s logic I/O must be set to
mode 8 or 9 via
HART/AMS so the sequencer and Oxymitter
4000 can communicate.
Depending on your system setup, a semi-
automatic calibration can be initiated by the
following methods:
(a) Oxymitter 4000. Press the CAL key on
the Oxymitter 4000 keypad.
(b) IMPS 4000. Use the IMPS 4000
keypad to change the InitCalX pa-
rameter of the CHANGE PRESETS
display mode from 0000 to 0001. Refer
to the IMPS 4000 Intelligent Multi-
probe Test Gas Sequencer Instruction
Bulletin for more information.
(c) HART. Use the HART Communicator
to access the O
2
CALIBRATE menu
and perform the O
2
CAL method. Refer
to paragraph 7-7 for the complete cali-
bration procedure.
(d) AMS. Refer to AMS documentation
for more information.
(e) Remote Contact. Initiate a calibration
from a remote location via the remote
contact input connection provided by an
IMPS 4000 or SPS 4000. Refer to the
documentation available for the control
system in use for more information.
Once a semi-automatic calibration is
initiated, by any of the methods previously
described, the Oxymitter 4000s
CALIBRATION RECOMMENDED alarm
signals an IMPS 4000 or SPS 4000 to initiate
a calibration. The sequencer sends an in cal
signal to the control room so that any auto-
matic control loops can be placed in manual.
Then, the sequencer begins to sequence the
calibration gases.
3. Manual Calibration. Manual calibrations
must be performed at the Oxymitter 4000 site
and require operator intervention throughout
the process.
Manual calibration instructions can also be
found, in condensed form, on the inside of
the right electronics housing cover (Figure
6-3).