  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 31
IB-106-340
xxvi
QUICK START GUIDE
FOR OXYMITTER 4000 SYSTEMS
Before using the Quick Start Guide, please read “WHAT YOU
NEED TO KNOW BEFORE INSTALLING AND WIRING A
ROSEMOUNT OXYMITTER 4000 OXYGEN TRANSMITTER” on
the preceding page.
1.
Install the Oxymitter 4000 in an appropriate location on the stack or duct. Refer to
Section II, paragraph 2-1.a for information on selecting a location for the
Oxymitter 4000.
2.
If using an SPS 4000, connect the calibration gasses to the appropriate fittings on
the SPS 4000 manifold.
3.
Connect reference air to the Oxymitter 4000 or SPS 4000, as applicable.
4.
If using an SPS 4000, make the following wire connections as shown in Figure 2:
line voltage, cal initiate-remote contact input, relay output, and 4-20 mA.
5.
If NOT using an SPS 4000, make the following wire connections as shown in
Figure 3: line voltage, 4-20 mA, and logic I/O.
6.
Verify the Oxymitter 4000 switch configuration is as desired. Refer to Section III,
paragraphs 3-1.c, 3-1.d, and 3-1.e.
7.
Apply power to the Oxymitter 4000; the cell heater will turn on. Allow
approximately one half hour for the cell to heat to operating temperature. Once the
ramp cycle has completed and the Oxymitter 4000 is at normal operation, proceed
with step 8.
8.
If using an SPS 4000, initiate a semi-automatic calibration.
9.
If NOT using an SPS 4000, perform a manual calibration. Refer to the QUICK
REFERENCE GUIDE manual calibration instructions on the following pages, or
Section VI, paragraph 6-2, Calibration, in this instruction bulletin.