  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 130
Instruction Manual
IM-106-340C, Rev. 4.4
July 2008
9-17
Hazardous Area Oxymitter 4000
Figure 9-8. Heater Strut
Assembly
8. See Figure 9-8. Grasp the wire loop and carefully slide the heater strut
assembly (32, Figure 9-3) out of the probe tube.
9. When replacing the strut, orient the probe so that the small calibration
gas tube lays at the 6 o'clock position of the probe tube. Align the slot on
the heater plate with the calibration gas line in the probe tube. Slide the
strut into the probe tube. It will turn to align the hole on the back plate of
the strut with the calibration gas line. When the hole and the calibration
gas line are aligned correctly, the strut will slide in the rest of the way.
10. As the strut installation nears completion, install the guide rod into the
calibration gas tube to assist in guiding the calibration gas tube through
the hole in the end of the strut.
11. Push down on the back plate of the strut to make sure you have spring
tension and then install the strut pressure clamp (34) on the back plate.
12. Install gas port fittings (29) in the CAL, REF, and VENT ports.
13. Replace the CAL and REF gas silicon tubes (34) and tube clamps (33).
14. Install the entire electronics per "Replacement of Entire Electronics (with
Housing)", steps 4 and 5.
15. Follow the instructions in "Probe Installation" in Section 2: Installation to
install the Hazardous Area Oxymitter 4000 into the stack or duct.
38860021
Wire
Loop
CeramicSupportRod
CellFlange
Heater
Ceramic
Diffuser
Assembly
V-Deflector