Model 6081-PWireless pH/ORP TransmitterInstruction ManualPN 51-6081P/rev.ANovember 2008
MODEL 6081-P pH/ORP TABLE OF CONTENTivThis page left blank intentionally
MODEL 6081-P pH/ORP SECTION 1.0DESCRIPTION AND SPECIFICATIONSSECTION 1.0DESCRIPTION AND SPECIFICATIONS1• HIGH ACCURACY AND RELIABILITY for monitoringa
MODEL 6081-P pH/ORP SECTION 1.0DESCRIPTION AND SPECIFICATIONS21.3 SPECIFICATIONS - WIRELESSOutput: WirelessHART V7Transmit Rate: User selectable, 1/se
3MODEL 6081-P pH/ORP SECTION 1.0DESCRIPTION AND SPECIFICATIONS1.5 Product CertificationsTelecommunication ComplianceAll wireless devices require certi
MODEL 6081-P pH/ORP SECTION 1.0DESCRIPTION AND SPECIFICATIONS1.7 ASSET MANAGEMENT SOLUTIONSAsset Management Solutions (AMS) is software that helps pla
MODEL 6081-P pH/ORP SECTION 2.0INSTALLATIONSECTION 2.0INSTALLATION2.1 Considerations2.2 Unpacking and Inspection2.3 Pre-Installation Set Up2.4 Mechani
62.3 PRE-INSTALLATION SETUP2.3.1 Temperature ElementThe Model 6081-P pH/ORP transmitter is compatible with sensors having Pt 100 and Pt 1000. Sensors
7WARNINGFailure to follow these installation guidelines could result in death or serious injury.• Make sure only qualified personnel perform the insta
8MODEL 6081-P pH/ORP SECTION 2.0INSTALLATION2.4.2 Mounting on a Flat Surface.Figure 2-1. Wall Mounting Installation Model 6081. Use Pipe/Wall Mountin
9MODEL 6081-P pH/ORP SECTION 2.0INSTALLATION2.4.3 Pipe Mounting.The pipe mounting kit (PN 23820-00/01) accommodates 1-1/2 to 2 in. pipe.Figure 2-2. Pi
ESSENTIAL INSTRUCTIONSREAD THIS PAGE BEFORE PROCEEDING!Rosemount Analytical designs, manufactures, and tests its products to meet many national and i
102.6 POWER MODULE INSTALLATION The section describes the procedure for installation of the power module (part number 00753-9220-0001). Thepower modu
4. With gentle pressure, insert the keyed connector on the power module into the receptacle (labeled Power Module Connection on the drawing). The p
12This page left blank intentionallyMODEL 6081-P pH/ORP SECTION 2.0INSTALLATION
MODEL 6081-P pH/ORP SECTION 3.0 WIRINGSECTION 3.0SENSOR WIRING3.1 GENERAL INFORMATIONpH and ORP sensors without preamps manufactured by Rosemount Anal
14MODEL 6081-P pH/ORP SECTION 3.0 WIRINGThis page left blank intentionally
15MODEL 6081-P pH/ORP SECTION 4.0INTRINSICALLY SAFE & EXPLOSION PROOF15SECTION 4.0INTRINSICALLY SAFE INSTALLATIONThis page left blank intentionall
16MODEL 6081-P pH/ORP SECTION 4.0INTRINSICALLY SAFE & EXPLOSION PROOFThis page left blank intentionally
17MODEL 6081-P pH/ORP SECTION 5.0 COMMISSIONINGSECTION 5.0COMMISSIONING5.1 Network Communications5.2 Device Network Configuration5.3 Verify Operation5
18MODEL 6081-P pH/ORP SECTION 5.0 OPERATION WITH REMOTE CONTROLLER5.3 VERIFY OPERATIONOperation can be verified in three locations, at the device via
19MODEL 6081-P pH/ORP SECTION 6.0 DISPLAY AND OPERATIONSECTION 6.0DISPLAY AND OPERATION6.1. DISPLAYThe Model 6081 has a two-linedisplay. Generally, t
About This DocumentThis manual contains instructions for installation and operation of the Model 6081-P Wireless pH/ORPTransmitter. The following list
206.3 MENUS - pHThe Model 6081 pH transmitter has four menus: CALIBRATE, PROGRAM, SIM PV, and DISPLAY. Under theCalibrate and Program menus are sever
21MODEL 6081-P pH/ORP SECTION 6.0 DISPLAY AND OPERATION6.5 SECURITY6.6.1 How the Security Code WorksUse security codes to prevent accidental or unwant
22MODEL 6081-P pH/ORP SECTION 6.0 DISPLAY AND OPERATIONThis page left blank intentionally22
23MODEL 6081-P pH/ORP SECTION 7.0OPERATION WITH MODEL 3757.1 Note on Model 375 HART Communicator7.2 Connecting the HART Communicator7.3 OperationSECT
247.3 Operation7.3.1 Off-line and On-line OperationThe Model 375 Communicator features off-line and on-line communications. On-line means the communic
25MODEL 6081-P pH/ORP SECTION 8.0PROGRAMMING THE TRANSMITTERSECTION 8.0PROGRAMMING THE TRANSMITTER 8.1 General8.2 Changing Start-up Settings8.3
26MODEL 6081-P pH/ORP SECTION 8.0PROGRAMMING THE TRANSMITTER8.3 CHOOSING AND CONFIGURING THE ANALYTICAL MEASUREMENT8.3.1 PurposeThis section describes
27MODEL 6081-P pH/ORP SECTION 8.0PROGRAMMING THE TRANSMITTER8.3.3 Procedure to configure: Measurement.To choose a menu item, move the cursor to the it
28MODEL 6081-P pH/ORP SECTION 8.0PROGRAMMING THE TRANSMITTER8.4 CHOOSING TEMPERATURE UNITS AND MANUAL OR AUTOMATIC TEMPERATURECOMPENSATION8.4.1 Purpo
29MODEL 6081-P pH/ORP SECTION 8.0PROGRAMMING THE TRANSMITTER8.5.2 Procedure: Setting a security code.1. Press MENU. The menu screen appears. Choose Pr
QUICK START GUIDEFOR MODEL 6081 WIRELESS pH TRANSMITTERD1. Refer to Section 2.0 Installation for installation instructions.2 . Wire the pH or ORP s
30MODEL 6081-P pH/ORP SECTION 8.0PROGRAMMING THE TRANSMITTER8.6 MAKING HART RELATED SETTINGSFor more information refer to Sec. 1.08.7 RESETTING FACTOR
31MODEL 6081-P pH/ORP SECTION 8.0PROGRAMMING THE TRANSMITTER8.8 SELECTING A DEFAULT SCREEN AND SCREEN CONTRAST8.8.1 PurposeThis section describes how
32This page left blank intentionallyMODEL 6081-P pH/ORP SECTION 8.0PROGRAMMING THE TRANSMITTER328.9 CHOOSING A DISPLAY TIMEOUT8.9.1 PurposeThe local t
33MODEL 6081-P pH/ORP SECTION 9.0CALIBRATION – TEMPERTURESECTION 9.0CALIBRATION — TEMPERATURE9.1 INTRODUCTIONThe Calibrate Menu allows the user to cal
34MODEL 6081-P pH/ORP SECTION 9.0CALIBRATION – TEMPERTURE9.2.2 Procedure1. Remove the sensor from the process. Place it in an insulated container of w
35MODEL 6081-P pH/ORP SECTION 10.0CALIBRATION – PHSECTION 10.0CALIBRATION — pH AND ORP10.1 INTRODUCTIONFor pH sensors, two-point buffer calibration i
36MODEL 6081-P pH/ORP SECTION 10.0CALIBRATION – PHFIGURE 10-1. Calibration Slope and OffsetDuring automatic calibration, the transmitter also measures
37MODEL 6081-P pH/ORP SECTION 10.0CALIBRATION – PH10.2 PROCEDURE — AUTO BUFFER CALIBRATION1. Obtain two buffer solutions. Ideally, the buffer values
3810.3 PROCEDURE — MANUAL TWO-POINT BUFFER CALIBRATION1. Obtain two buffer solutions. Ideally, the buffer values should bracket the range of pH valu
39MODEL 6081-P pH/ORP SECTION 10.0CALIBRATION – PH10.4 PROCEDURE — STANDARDIZATION1. The pH measured by the transmitter can be changed to match the
EQuick StartMenu English Francais Espanol >>Deutsch Italiano Portuguese >>Measure? pH Redox
4010.5 PROCEDURE — ENTERING A KNOWN SLOPE VALUE.1. If the electrode slope is known from other measurements, it can be entered directly into the tran
41MODEL 6081-P pH/ORP SECTION 10.0CALIBRATION – PH10.6 ORP CALIBRATION10.6.1 Purpose1. For process control, it is often important to make the measured
This page left blank intentionally42MODEL 6081-P pH/ORP SECTION 10.0CALIBRATION – PH
43MODEL 6081-P pH/ORP SECTION 11.0MAINTENANCESECTION 11.0MAINTENANCE11.1 OVERVIEWThis section gives general procedures for routine maintenance of the
44MODEL 6081-P pH/ORP SECTION 11.0MAINTENANCE11.3 pH SENSOR MAINTENANCE11.3.1 Frequency of CleaningThe frequency at which a sensor should be inspect
4511.3.3 Checking the Reference Electrode.Some processes contain substances, for example, sulfides, that poison the reference electrode. Poisoning al
4611.4 ORP SENSOR MAINTENANCE11.4.1 Frequency of CleaningThe frequency at which an ORP sensor should be inspected and cleaned can be determined onl
4711.6 POWER MODULE REPLACEMENTExpected power module (part number 00753-9220-0001) life is two years at reference conditions. This sectiondescribes t
48This page left blank intentionally48MODEL 6081-P pH/ORP SECTION 11.0MAINTENANCE11.6.1 Handling ConsiderationsThe power module with the wireless unit
49MODEL 6081-P pH/ORP SECTION 12.0 RETURN OF MATERIALSSECTION 12.0RETURN OF MATERIAL12.1 GENERAL.To expedite the repair and return of instruments, pro
FFIGURE 1-1. MENU TREE FOR MODEL 6081 pH Wireless TRANSMITTER
50This page left blank intentionallyMODEL 6081-P pH/ORP SECTION 12.0 RETURN OF MATERIALS
51WARRANTYGoods and part(s) (excluding consumables) manufactured by Seller are warranted to be free from defects in workman-ship and material under no
Credit Cards for U.S. Purchases Only.The right people,the right answers,right now.ON-LINE ORDERING NOW AVAILABLE ON OUR WEB SITEhttp://www.raihome.com
MODEL 6081-P pH/ORP TABLE OF CONTENTSMODEL 6081-P pH/ORPWIRELESS TRANSMITTERTABLE OF CONTENTSSection Title Page1.0 DESCRIPTION AND SPECIFICATIONS ...
MODEL 6081-P pH/ORP TABLE OF CONTENTSTABLE OF CONTENTS CONT’Dii8.0 PROGRAMMING THE TRANSMITTER.....