Model 5081-T
Two-Wire Toroidal Conductivity Transmitter
Instruction Manual
PN 51-5081T/rev.D
February 2006
Page view 0
1 2 3 4 5 6 ... 97 98

Summary of Contents

Page 1 - Model 5081-T

Model 5081-TTwo-Wire Toroidal Conductivity TransmitterInstruction ManualPN 51-5081T/rev.DFebruary 2006

Page 2 - INSTRUCTIONS

4MODEL 5081-T SECTION 1.0DESCRIPTION AND SPECIFICATIONS1.4 TRANSMITTER DISPLAY DURINGCALIBRATION AND PROGRAMMING(FIGURE 1-1)1. Continuous display of

Page 3 - TWO-WIRE TRANSMITTER

5MODEL 5081-T SECTION 1.0DESCRIPTION AND SPECIFICATIONS1.6 HART COMMUNICATIONS1.6.1 OVERVIEW OF HART COMMUNICATIONHART (highway addressable remote tr

Page 4 - TABLE OF CONTENTS CONT’D

6MODEL 5081-T SECTION 1.0DESCRIPTION AND SPECIFICATIONS1.8 ASSET MANAGEMENT SOLUTIONSAsset Management Solutions (AMS) is software that helps plant per

Page 5 - LIST OF FIGURES

MODEL 5081-T SECTION 1.0DESCRIPTION AND SPECIFICATIONSFIGURE 1-5. AMS MAIN MENU TOOLS7

Page 6 - LIST OF TABLES

8MODEL 5081-T SECTION 2.0INSTALLATIONSECTION 2.0INSTALLATION2.1 Unpacking and Inspection2.2 Orienting the Display Board2.3 Mechanical Installation2.4

Page 7 - SECTION 1.0

9MODEL 5081-T SECTION 2.0INSTALLATIONFIGURE 2-1. Mounting the Model 5081-T Toroidal Conductivity Transmitter on a Flat SurfaceMILLIMETERINCH2.3.2 Moun

Page 8 - 1.2 SPECIFICATIONS

10MODEL 5081-T SECTION 2.0INSTALLATIONFIGURE 2-2. Using the Pipe Mounting Kit to Attach the Model 5081-T Conductivity Transmitter to a PipeMILLIMETERI

Page 9 - RECOMMENDED SENSORS:

11MODEL 5081-T SECTION 2.0INSTALLATIONTABLE 2-1. Model 5081-T Sensor Selection2.3.4 Inductive Loops. The Model 5081-T conductivity transmitter is de

Page 10 - #"c""

12MODEL 5081-T SECTION 2.0INSTALLATION2.4 POWER SUPPLY/CURRENT LOOP— MODEL 5081-T-HT2.4.1 Power Supply and Load Requirements.Refer to Figure 2-3.The

Page 11 - 1.6 HART COMMUNICATIONS

13MODEL 5081-T SECTION 2.0INSTALLATION2.5 POWER SUPPLY WIRING FOR MODEL 5081-T-FF/FI2.5.1 Power Supply Wiring. Refer to Figure 2-5 andFigure 2-6.Run

Page 12 - 1.7 FOUNDATION FIELDBUS

Emerson Process ManagementRosemount Analytical Inc.2400 Barranca ParkwayIrvine, CA 92606 USATel: (949) 757-8500Fax: (949) 474-7250http://www.raihome.

Page 13 - MODEL 5081-T SECTION 1.0

14MODEL 5081-T SECTION 3.0 WIRINGSECTION 3.0WIRING3.1 Sensor Wiring3.2 Electrical Installation3.1.1 WIRING THROUGH A JUNCTION BOXThe sensor can be w

Page 14 - INSTALLATION

15MODEL 5081-T SECTION 3.0 WIRING3.1.2 POWER WIRING MODEL 5081-T-HTFor general purpose areas, wire power as shownin Figure 3-2. For hazardous areas, p

Page 15 - MILLIMETER

16MODEL 5081-T SECTION 3.0 WIRING3.2 ELECTRICAL INSTALLATIONAll Rosemount Analytical contacting conductivity sensors with PT100 RTD or PT1000 RTD are

Page 16 - DWG. NO. REV

MODEL 5081-T SECTION 3.0 WIRINGFIGURE 3-6. Wiring Models 222, 225, 226, 228, 242, & 247 sensors to Model 5081-T transmitter17

Page 17

18MODEL 5081-T SECTION 3.0 WIRINGFIGURE 3-7. Wiring Models 222, 225, 226, & 228 sensors to Model 5081-T transmitter

Page 18 - — MODEL 5081-T-HT

MODEL 5081-T SECTION 4.0INTRINSICALLY SAFE & EXPLOSION PROOFSECTION 4.0INTRINSICALLY SAFE & EXPLOSION PROOF19FIGURE 4-1. Model 5081-T-HT Infra

Page 19 - Transmitter

20MODEL 5081-T SECTION 4.0INTRINSICALLY SAFE & EXPLOSION PROOFFIGURE 4-3. FM Explosion-Proof Installation for Model 5081-T-HT4.1 INTRINSICALLY SAF

Page 20 - SECTION 3.0

MODEL 5081-T SECTION 4.0INTRINSICALLY SAFE & EXPLOSION PROOF21FIGURE 4-4. FM Intrinsically Safe Installation for Model 5081-T-HT

Page 21 - MODEL 5081-T SECTION 3.0

22MODEL 5081-T SECTION 4.0INTRINSICALLY SAFE & EXPLOSION PROOFFIGURE 4-5. CSA Intrinsically Safe Installation for Model 5081-T-HT

Page 22 - 3.2 ELECTRICAL INSTALLATION

MODEL 5081-T SECTION 4.0INTRINSICALLY SAFE & EXPLOSION PROOF23FIGURE 4-6. ATEX Intrisically Safe Label for Model 5081-T-HT

Page 23

iMODEL 5081-T TABLE OF CONTENTSMODEL 5081-TTWO-WIRE TRANSMITTERTABLE OF CONTENTSSection Title Page1.0 DESCRIPTION AND SPECIFICATIONS ...

Page 24

24MODEL 5081-T SECTION 4.0INTRINSICALLY SAFE & EXPLOSION PROOFFIGURE 4-7. ATEX Intrisically Safe Label for Model 5081-T-HT

Page 25 - IRC - INFRARED REMOTE CONTROL

MODEL 5081-T SECTION 4.0INTRINSICALLY SAFE & EXPLOSION PROOF25FIGURE 4-8. FM Explosion-Proof Installation for Model 5081-T-FF4.2 INTRINSICALLY SAF

Page 26 - MODEL 5081-T SECTION 4.0

26MODEL 5081-T SECTION 4.0INTRINSICALLY SAFE & EXPLOSION PROOFFIGURE 4-9. FM Intrinsically Safe Installation for Model 5081-T-FF

Page 27

MODEL 5081-T SECTION 4.0INTRINSICALLY SAFE & EXPLOSION PROOF27FIGURE 4-10. CSA Intrinsically Safe Installation for Model 5081-T-FF

Page 28

28MODEL 5081-T SECTION 4.0INTRINSICALLY SAFE & EXPLOSION PROOFFIGURE 4-11. ATEX Intrisically Safe Label for Model 5081-T-FF

Page 29

MODEL 5081-T SECTION 4.0INTRINSICALLY SAFE & EXPLOSION PROOF29FIGURE 4-12. ATEX Intrinsically Safe Installation for Model 5081-T-FF

Page 30

30FIGURE 4-13. FM Explosion-Proof Installation for Model 5081-T-FI5/6/04J. FLOCK5/6/04J. FLOCK 70°C MAXCLASS III, DIV 1CLASS II, DIV 1, GPS E-GCL

Page 31

31FM A5/6/04J. FLOCK5/6/04J. FLOCK.140±.0051.30±.02.650±.015.1252.180±.0052.56±.02.120±.015Ø .125R .255-6-04

Page 32

32FM BLa, Lt OR Lo; OR Lc/Rc (La/Ra OR Lo/Ro) AND Li/Ri (La/Ra OR Lo/Ro)1. INSTALLATION SHOULD BE IN ACCORDANCE WITH ANSI/

Page 33

33THIS DOCUMENT ISCERTIFIED BYREVREVREVREVREVREVREVISIONS NOT PERMITTEDW/O AGENCY APPROVALDIMENSIONS ARE IN INCHESREMOVE BURRS & SHARP EDGES .020

Page 34

MODEL 5081-T TABLE OF CONTENTSTABLE OF CONTENTS CONT’Dii7.0 PROGRAMMING...

Page 35

Rosemount Analytical,Uniloc Divisi on2400 Barranca PkwyIrvine, CA 92 606UNLESS OTHERWISE SPECIFIEDDIMENSION S ARE IN INCHESREMOVE BURRS & SHARP E

Page 36 - SAFE AREA SAFE AREA

359241514-01/ADIMENSIONS ARE IN INCHESREMOVE BURRS & SHARP EDGES .020 MAXMACHINED FILLET RADII .020 MAX-ITEMPART NODESCRIPTIONQTYSCALESIZEREVDWG N

Page 37 - 9241515-01

Rosemount Analytical,Uniloc Division2400 Bar ranca Pk wyIrvine, CA 92606UNLESS OTHERWISE SPECIFIEDDIMENSIONS ARE IN INCHESREMOVE BURRS & SHARP EDG

Page 38

37MODEL 5081-T SECTION 5.0 DISPLAY AND OPERATIONSECTION 5.0DISPLAY AND OPERATION5.1 Displays5.2 Infrared Remote Controller (IRC) - Key Functions5.3 Qu

Page 39 - 9241516-01

385.2 INFRARED REMOTE CONTROLLER (IRC) - KEY FUNCTIONSThe infrared remote controller is used to calibrate and program the transmitter and to read diag

Page 40

39MODEL 5081-T SECTION 5.0 DISPLAY AND OPERATION5.3 QUICK START FOR MODEL 5081-T-HT (HART)1. On the Remote, press PROG, NEXT, NEXT, ENTER.2. Use th

Page 41 - 9241514 01

MODEL 5081-T SECTION 5.0 DISPLAY AND OPERATION5.4 QUICK START FOR MODEL 5081-T-FF/FI (FOUNDATION FIELDBUS)1. On the Remote, press PROG, NEXT, ENTER.

Page 42

41MODEL 5081-T SECTION 5.0 DISPLAY AND OPERATION5.5 MENU TREE - ConductivityThe Model 5081-T transmitter has three menus: CALIBRATE, PROGRAM, and DIA

Page 43

MODEL 5081-T SECTION 5.0 DISPLAY AND OPERATION42PROGRAM MENU MNEMONICS9>U:>UCurrent output menu header&7-4mA current output (setpoint)$"

Page 44 - DISPLAY AND OPERATION

43MODEL 5081-T SECTION 5.0 DISPLAY AND OPERATION5.6 DIAGNOSTIC MESSAGESWhenever a warning or fault limit has been exceeded, the transmitter displays

Page 45

iiiMODEL 5081-T TABLE OF CONTENTSLIST OF FIGURESNumber Title Page1-1 Transmitter Display During Calibration and Programming ...

Page 46 - FOUNDATION FIELDBUS)

44MODEL 5081-T SECTION 5.0 DISPLAY AND OPERATIONVARIABLE NAME MNEMONIC FACTORY SETTINGS CUSTOMER SETTINGSProgram MenuOutput9VURVU– ___________4 mA &am

Page 47 - '"""

455.8 SECURITY5.8.1 General. Use the programmable security code to protect program and calibrationsettings from accidentally being changed. The trans

Page 48

46MODEL 5081-T SECTION 6.0START-UP AND CALIBRATIONSECTION 6.0START-UP AND CALIBRATION6.1 ACCESSING THE CALIBRATE MENUThe “Calibrate” menu is used to c

Page 49 - 5.7 DEFAULT SETTINGS

476.2 CALIBRATE MENU To access the “Calibrate” menu, press the CALkey on the Infrared Remote Control. If security hasbeen enabled, the secondary proce

Page 50

48MODEL 5081-T SECTION 6.0START-UP AND CALIBRATION6.2.5 Temp Slope1. Press NEXT to enter the tEMP SLOPE menu.The correct temperature slope must be ent

Page 51 - 5.9 USING HOLD

49MODEL 5081-T SECTION 6.0START-UP AND CALIBRATIONTABLE 6-1. CALIBRATE MENU MNEMONICS/-[5ES-U1 Calibrate menu header/-[ Sensor calibrationU17:!<[9:

Page 52 - START-UP AND CALIBRATION

50MODEL 5081-T SECTION 7.0PROGRAMMINGSECTION 7.0PROGRAMING7.1 General7.2 Output7.3 Temp7.4 Display7.5 HART7.6 Setup Cust7.7 Range7.8 Default7.1 GENERA

Page 53 - 6.2 CALIBRATE MENU

MODEL 5081-T SECTION 7.0PROGRAMMING7.2 OUTPUT (5081-HT only)7.2.1 PurposeThis section describes how to do the following:1. assign values to the 4 and

Page 54 - CALIBRATE

52MODEL 5081-T SECTION 7.0PROGRAMMING7.2.3 Procedure1. Press PROG on the remote controller. The OutPut submenu appears.2. Press ENTER. The screen disp

Page 55

MODEL 5081-T SECTION 7.0PROGRAMMING7.3 TEMP7.3.1 PurposeThis section describes how to do the following:1. Enable and disable automatic temperature co

Page 56 - PROGRAMING

ivMODEL 5081-T TABLE OF CONTENTSLIST OF FIGURES - CONT’DNumber Title Page5-1 Process Display Screen ...

Page 57 - 7.2 OUTPUT (5081-HT only)

54MODEL 5081-T SECTION 7.0PROGRAMMING1. Press PROG on the remote controller.2. Press NEXT until the diSPLAy submenu appears. Press ENTER.3. Press  or

Page 58

7.5 HARTIn multi-drop operation, polling addresses can be more conveniently set and debugging more conveniently performedusing the infrared remote con

Page 59 - PROGRAMMING

567.6 SETUP CUSTThe Model 5081-T contains a curve fitting program that can create a second order curve for3 to 5 user supplied data points. If only tw

Page 60 - 7.4 DISPLAY

MODEL 5081-T SECTION 7.0PROGRAMMING7.7 RANGE7.7.1 PurposeThis section provides the steps to select automatic ranging or a specific fixed range of meas

Page 61 - 7.5 HART

58MODEL 5081-T SECTION 8.0FOUNDATION FIELDBUS OPERATIONSECTION 8.0FOUNDATION FIELDBUS OPERATIONThis section covers basic transmitter operation and sof

Page 62 - EXIT NEXT ENTER

MODEL 5081-T SECTION 9.0 OPERATION WITH MODEL 375SECTION 9.0OPERATION WITH MODEL 3759.1 Note on Model 375 or 275 CommunicatorThe Model 375 or 275 Comm

Page 63 - OUNDATION Fieldbus hosts or

609.3 Operation9.3.1 Off-line and On-line OperationThe Model 375 Communicator features off-line and on-line communications. On-line means the communic

Page 64 - FOUNDATION fieldbus com

MODEL 5081-T SECTION 9.0 OPERATION WITH MODEL 375--------------------------------------------------------------------------------5081-C/T 275, 375 Men

Page 65 - OPERATION WITH MODEL 375

62MODEL 5081-T SECTION 9.0 OPERATION WITH MODEL 375Basic setup (continued)Device informationDistributorModelDev idTagDatePhysicl signl codeWrite prote

Page 66 - 9.3 Operation

MODEL 5081-T SECTION 9.0 OPERATION WITH MODEL 375Diag override (continued)RTD fail [ON, OFF]Sense line open [ON, OFF]Temp high/low [ON, OFF]Signal con

Page 67 - Soln offset **

1MODEL 5081-T SECTION 1.0DESCRIPTION AND SPECIFICATIONSSECTION 1.0DESCRIPTION AND SPECIFICATIONS• CHOICE OF COMMUNICATION PROTOCOL: HART or FOUNDATION

Page 68 - Zero warning [ON, OFF]

64MODEL 5081-T SECTION 9.0 OPERATION WITH MODEL 375Local DisplayAO LOI units [mA, %]Xmtr IDLoad Default Conf.ReviewPVPV AOPV LRVPV URV----------------

Page 69 - MODEL 5081-T SECTION 9.0

MODEL 5081-T SECTION 9.0 OPERATION WITH MODEL 3755081-C/T-FF/FI 375 Menu Tree-----------------------RESOURCEIdentificationMANUFACT_IDDEV_TYPEDEV_REVDD

Page 70

66MODEL 5081-T SECTION 9.0 OPERATION WITH MODEL 375Transducer ErrorST_REVBLOCK_ERRAdditional transmitter statusFault history 0Fault history 1Fault his

Page 71

MODEL 5081-T SECTION 9.0 OPERATION WITH MODEL 375ALERT_KEYHI_HI_LIMHI_HI_PRIHI_LIMHI_PRIIO_OPTSL_TYPELO_LO_LIMLO_LO_PRILO_LIMLO_PRIMODE_BLK: TargetMOD

Page 72 - Snsr Cal Meth

68MODEL 5081-T SECTION 9.0 OPERATION WITH MODEL 375SIMULATE: Transducer StatusSIMULATE: Transducer ValueSIMULATE: Simulate En/DisableXD_SCALE: EU at 1

Page 73

MODEL 5081-T SECTION 9.0 OPERATION WITH MODEL 375OUT_SCALE: EU at 0%OUT_SCALE: Units IndexOUT_SCALE: DecimalPV_SCALE: EU at 100%PV_SCALE: EU at 0%PV_S

Page 74

70MODEL 5081-T SECTION 9.0 OPERATION WITH MODEL 375OnlineBK_CAL_IN: StatusBK_CAL_IN: ValueBK_CAL_OUT: StatusBK_CAL_OUT: ValueBLOCK_ERRBYPASSCAS_IN: St

Page 75

MODEL 5081-T SECTION 9.0 OPERATION WITH MODEL 375DV_HI_ALM: Time StampDV_HI_ALM: SubcodeDV_HI_ALM: Float ValueDV_LO_ALM: UnacknowledgedDV_LO_ALM: Alar

Page 76

72MODEL 5081-T SECTION 9.0 OPERATION WITH MODEL 375TRK_SCALE: Units IndexTRK_SCALE: DecimalTRK_IN_D: StatusTRK_IN_D: ValueTRK_VAL: StatusTRK_VAL: Valu

Page 77

MODEL 5081-T SECTION 9.0 OPERATION WITH MODEL 375StdDevCap StdDevSchedulingDetailPhysical Device TagAddressDevice IDDevice RevisionAdvancedStack Capab

Page 78

2MODEL 5081-T SECTION 1.0DESCRIPTION AND SPECIFICATIONS1.2 SPECIFICATIONS1.2.1 GENERAL SPECIFICATIONSEnclosure: Cast aluminum containing less than 6%

Page 79

74MODEL 5081-T SECTION 9.0 OPERATION WITH MODEL 375Normal Rcv SAP LasTime Available SAP Src SinkNormal SAP Src SinkUrgent SAP Src SinkTime Available R

Page 80

75MODEL 5081-T SECTION 10.0DIAGNOSIS AND TROUBLESHOOTINGSECTION 10.0DIAGNOSIS AND TROUBLESHOOTING10.1 OVERVIEWThe Model 5081-T transmitters automatica

Page 81 - DIAGNOSIS AND TROUBLESHOOTING

76MODEL 5081-T SECTION 10.0DIAGNOSIS AND TROUBLESHOOTING-E< Absolute conductivity (µS/cm or mS/cm)"!DLS Sensor zero in air/1[[!/QPTU Sensor ce

Page 82 - MODEL 5081-T SECTION 10.0

MODEL 5081-T SECTION 10.0DIAGNOSIS AND TROUBLESHOOTING10.2 FAULT CONDITIONSThree classes of error conditions/problems are detectedand differentiated b

Page 83

78MODEL 5081-T SECTION 10.0DIAGNOSIS AND TROUBLESHOOTING10.3 DIAGNOSTIC MESSAGESThe Model 5081-T transmitter’s diagnostics constantly monitor the con

Page 84 - 10.3 DIAGNOSTIC MESSAGES

79MODEL 5081-T SECTION 10.0DIAGNOSIS AND TROUBLESHOOTINGSYMPTOM ACTIONWrong temperature reading. Perform a temperature standardization. Verify sensor&

Page 85

80MODEL 5081-T SECTION 10.0DIAGNOSIS AND TROUBLESHOOTING10.5 SYSTEMATIC TROUBLESHOOTINGIf the Quick Troubleshooting Guide does not resolve the error,

Page 86

81MODEL 5081-T SECTION 10.0DIAGNOSIS AND TROUBLESHOOTING10.6 RTD RESISTANCE VALUESTable 10-4 is a ready reference of RTD resistance valuesat various t

Page 87

82MODEL 5081-T SECTION 10.0DIAGNOSIS AND TROUBLESHOOTING10.7 WARNING AND FAULT MESSAGESThe Model 5081-T transmitter continuously monitors the sensor a

Page 88

MODEL 5081-T SECTION 10.0DIAGNOSIS AND TROUBLESHOOTINGMessage Explanation See SectionOuEr rAnGE Over range, measurement exceeds display limit 10.8.1SE

Page 89

MODEL 5081-T SECTION 1.0DESCRIPTION AND SPECIFICATIONS1.2.3 TRANSMITTER SPECIFICATIONS @ 25°CMeasured Range*: 50 to 2,000,000 µS/cm (see chart)Accura

Page 90

84MODEL 5081-T SECTION 10.0DIAGNOSIS AND TROUBLESHOOTING10.8.3 CAL ErrorAt the end of the calibration step, the transmitter calculates the sensitivit

Page 91

MODEL 5081-T SECTION 10.0DIAGNOSIS AND TROUBLESHOOTING10.8.7 OFFSEt ErrThe OFFSEt Err message appears if the zero offset (in mV) exceeds the programm

Page 92

86MODEL 5081-T SECTION 11.0MAINTENANCESECTION 11.0MAINTENANCE11.3 TRANSMITTER MAINTENANCEPeriodically clean the transmitter window with house-hold amm

Page 93 - MAINTENANCE

87MODEL 5081-T SECTION 11.0MAINTENANCEFIGURE 11-2. Exploded View of Model 5081-T TransmitterThree screws (part 13 in the drawing) hold the three circu

Page 94 - THEORY OF OPERATION

8812.1 OVERVIEWThis section is a general description of how the Model5081-T Transmitter operates. This section is for thoseusers who desire a greater

Page 95 - RETURN OF MATERIAL

MODEL 5081-T SECTION 13.0RETURN OF MATERIALSECTION 13.0RETURN OF MATERIAL13.1 GENERAL.To expedite the repair and return of instruments, propercommunic

Page 96

90This page intentionally left blank.

Page 97 - Irvine, CA 92606

WARRANTYGoods and part(s) (excluding consumables) manufactured by Seller are warranted to be free from defects in workman-ship and material under norm