4MODEL 5081-T SECTION 1.0DESCRIPTION AND SPECIFICATIONS1.4 TRANSMITTER DISPLAY DURINGCALIBRATION AND PROGRAMMING(FIGURE 1-1)1. Continuous display of
5MODEL 5081-T SECTION 1.0DESCRIPTION AND SPECIFICATIONS1.6 HART COMMUNICATIONS1.6.1 OVERVIEW OF HART COMMUNICATIONHART (highway addressable remote tr
6MODEL 5081-T SECTION 1.0DESCRIPTION AND SPECIFICATIONS1.8 ASSET MANAGEMENT SOLUTIONSAsset Management Solutions (AMS) is software that helps plant per
MODEL 5081-T SECTION 1.0DESCRIPTION AND SPECIFICATIONSFIGURE 1-5. AMS MAIN MENU TOOLS7
8MODEL 5081-T SECTION 2.0INSTALLATIONSECTION 2.0INSTALLATION2.1 Unpacking and Inspection2.2 Orienting the Display Board2.3 Mechanical Installation2.4
9MODEL 5081-T SECTION 2.0INSTALLATIONFIGURE 2-1. Mounting the Model 5081-T Toroidal Conductivity Transmitter on a Flat SurfaceMILLIMETERINCH2.3.2 Moun
10MODEL 5081-T SECTION 2.0INSTALLATIONFIGURE 2-2. Using the Pipe Mounting Kit to Attach the Model 5081-T Conductivity Transmitter to a PipeMILLIMETERI
11MODEL 5081-T SECTION 2.0INSTALLATIONTABLE 2-1. Model 5081-T Sensor Selection2.3.4 Inductive Loops. The Model 5081-T conductivity transmitter is de
12MODEL 5081-T SECTION 2.0INSTALLATION2.4 POWER SUPPLY/CURRENT LOOP— MODEL 5081-T-HT2.4.1 Power Supply and Load Requirements.Refer to Figure 2-3.The
13MODEL 5081-T SECTION 2.0INSTALLATION2.5 POWER SUPPLY WIRING FOR MODEL 5081-T-FF/FI2.5.1 Power Supply Wiring. Refer to Figure 2-5 andFigure 2-6.Run
Emerson Process ManagementRosemount Analytical Inc.2400 Barranca ParkwayIrvine, CA 92606 USATel: (949) 757-8500Fax: (949) 474-7250http://www.raihome.
14MODEL 5081-T SECTION 3.0 WIRINGSECTION 3.0WIRING3.1 Sensor Wiring3.2 Electrical Installation3.1.1 WIRING THROUGH A JUNCTION BOXThe sensor can be w
15MODEL 5081-T SECTION 3.0 WIRING3.1.2 POWER WIRING MODEL 5081-T-HTFor general purpose areas, wire power as shownin Figure 3-2. For hazardous areas, p
16MODEL 5081-T SECTION 3.0 WIRING3.2 ELECTRICAL INSTALLATIONAll Rosemount Analytical contacting conductivity sensors with PT100 RTD or PT1000 RTD are
MODEL 5081-T SECTION 3.0 WIRINGFIGURE 3-6. Wiring Models 222, 225, 226, 228, 242, & 247 sensors to Model 5081-T transmitter17
18MODEL 5081-T SECTION 3.0 WIRINGFIGURE 3-7. Wiring Models 222, 225, 226, & 228 sensors to Model 5081-T transmitter
MODEL 5081-T SECTION 4.0INTRINSICALLY SAFE & EXPLOSION PROOFSECTION 4.0INTRINSICALLY SAFE & EXPLOSION PROOF19FIGURE 4-1. Model 5081-T-HT Infra
20MODEL 5081-T SECTION 4.0INTRINSICALLY SAFE & EXPLOSION PROOFFIGURE 4-3. FM Explosion-Proof Installation for Model 5081-T-HT4.1 INTRINSICALLY SAF
MODEL 5081-T SECTION 4.0INTRINSICALLY SAFE & EXPLOSION PROOF21FIGURE 4-4. FM Intrinsically Safe Installation for Model 5081-T-HT
22MODEL 5081-T SECTION 4.0INTRINSICALLY SAFE & EXPLOSION PROOFFIGURE 4-5. CSA Intrinsically Safe Installation for Model 5081-T-HT
MODEL 5081-T SECTION 4.0INTRINSICALLY SAFE & EXPLOSION PROOF23FIGURE 4-6. ATEX Intrisically Safe Label for Model 5081-T-HT
iMODEL 5081-T TABLE OF CONTENTSMODEL 5081-TTWO-WIRE TRANSMITTERTABLE OF CONTENTSSection Title Page1.0 DESCRIPTION AND SPECIFICATIONS ...
24MODEL 5081-T SECTION 4.0INTRINSICALLY SAFE & EXPLOSION PROOFFIGURE 4-7. ATEX Intrisically Safe Label for Model 5081-T-HT
MODEL 5081-T SECTION 4.0INTRINSICALLY SAFE & EXPLOSION PROOF25FIGURE 4-8. FM Explosion-Proof Installation for Model 5081-T-FF4.2 INTRINSICALLY SAF
26MODEL 5081-T SECTION 4.0INTRINSICALLY SAFE & EXPLOSION PROOFFIGURE 4-9. FM Intrinsically Safe Installation for Model 5081-T-FF
MODEL 5081-T SECTION 4.0INTRINSICALLY SAFE & EXPLOSION PROOF27FIGURE 4-10. CSA Intrinsically Safe Installation for Model 5081-T-FF
28MODEL 5081-T SECTION 4.0INTRINSICALLY SAFE & EXPLOSION PROOFFIGURE 4-11. ATEX Intrisically Safe Label for Model 5081-T-FF
MODEL 5081-T SECTION 4.0INTRINSICALLY SAFE & EXPLOSION PROOF29FIGURE 4-12. ATEX Intrinsically Safe Installation for Model 5081-T-FF
30FIGURE 4-13. FM Explosion-Proof Installation for Model 5081-T-FI5/6/04J. FLOCK5/6/04J. FLOCK 70°C MAXCLASS III, DIV 1CLASS II, DIV 1, GPS E-GCL
31FM A5/6/04J. FLOCK5/6/04J. FLOCK.140±.0051.30±.02.650±.015.1252.180±.0052.56±.02.120±.015Ø .125R .255-6-04
32FM BLa, Lt OR Lo; OR Lc/Rc (La/Ra OR Lo/Ro) AND Li/Ri (La/Ra OR Lo/Ro)1. INSTALLATION SHOULD BE IN ACCORDANCE WITH ANSI/
33THIS DOCUMENT ISCERTIFIED BYREVREVREVREVREVREVREVISIONS NOT PERMITTEDW/O AGENCY APPROVALDIMENSIONS ARE IN INCHESREMOVE BURRS & SHARP EDGES .020
MODEL 5081-T TABLE OF CONTENTSTABLE OF CONTENTS CONT’Dii7.0 PROGRAMMING...
Rosemount Analytical,Uniloc Divisi on2400 Barranca PkwyIrvine, CA 92 606UNLESS OTHERWISE SPECIFIEDDIMENSION S ARE IN INCHESREMOVE BURRS & SHARP E
359241514-01/ADIMENSIONS ARE IN INCHESREMOVE BURRS & SHARP EDGES .020 MAXMACHINED FILLET RADII .020 MAX-ITEMPART NODESCRIPTIONQTYSCALESIZEREVDWG N
Rosemount Analytical,Uniloc Division2400 Bar ranca Pk wyIrvine, CA 92606UNLESS OTHERWISE SPECIFIEDDIMENSIONS ARE IN INCHESREMOVE BURRS & SHARP EDG
37MODEL 5081-T SECTION 5.0 DISPLAY AND OPERATIONSECTION 5.0DISPLAY AND OPERATION5.1 Displays5.2 Infrared Remote Controller (IRC) - Key Functions5.3 Qu
385.2 INFRARED REMOTE CONTROLLER (IRC) - KEY FUNCTIONSThe infrared remote controller is used to calibrate and program the transmitter and to read diag
39MODEL 5081-T SECTION 5.0 DISPLAY AND OPERATION5.3 QUICK START FOR MODEL 5081-T-HT (HART)1. On the Remote, press PROG, NEXT, NEXT, ENTER.2. Use th
MODEL 5081-T SECTION 5.0 DISPLAY AND OPERATION5.4 QUICK START FOR MODEL 5081-T-FF/FI (FOUNDATION FIELDBUS)1. On the Remote, press PROG, NEXT, ENTER.
41MODEL 5081-T SECTION 5.0 DISPLAY AND OPERATION5.5 MENU TREE - ConductivityThe Model 5081-T transmitter has three menus: CALIBRATE, PROGRAM, and DIA
MODEL 5081-T SECTION 5.0 DISPLAY AND OPERATION42PROGRAM MENU MNEMONICS9>U:>UCurrent output menu header&7-4mA current output (setpoint)$"
43MODEL 5081-T SECTION 5.0 DISPLAY AND OPERATION5.6 DIAGNOSTIC MESSAGESWhenever a warning or fault limit has been exceeded, the transmitter displays
iiiMODEL 5081-T TABLE OF CONTENTSLIST OF FIGURESNumber Title Page1-1 Transmitter Display During Calibration and Programming ...
44MODEL 5081-T SECTION 5.0 DISPLAY AND OPERATIONVARIABLE NAME MNEMONIC FACTORY SETTINGS CUSTOMER SETTINGSProgram MenuOutput9VURVU– ___________4 mA &am
455.8 SECURITY5.8.1 General. Use the programmable security code to protect program and calibrationsettings from accidentally being changed. The trans
46MODEL 5081-T SECTION 6.0START-UP AND CALIBRATIONSECTION 6.0START-UP AND CALIBRATION6.1 ACCESSING THE CALIBRATE MENUThe “Calibrate” menu is used to c
476.2 CALIBRATE MENU To access the “Calibrate” menu, press the CALkey on the Infrared Remote Control. If security hasbeen enabled, the secondary proce
48MODEL 5081-T SECTION 6.0START-UP AND CALIBRATION6.2.5 Temp Slope1. Press NEXT to enter the tEMP SLOPE menu.The correct temperature slope must be ent
49MODEL 5081-T SECTION 6.0START-UP AND CALIBRATIONTABLE 6-1. CALIBRATE MENU MNEMONICS/-[5ES-U1 Calibrate menu header/-[ Sensor calibrationU17:!<[9:
50MODEL 5081-T SECTION 7.0PROGRAMMINGSECTION 7.0PROGRAMING7.1 General7.2 Output7.3 Temp7.4 Display7.5 HART7.6 Setup Cust7.7 Range7.8 Default7.1 GENERA
MODEL 5081-T SECTION 7.0PROGRAMMING7.2 OUTPUT (5081-HT only)7.2.1 PurposeThis section describes how to do the following:1. assign values to the 4 and
52MODEL 5081-T SECTION 7.0PROGRAMMING7.2.3 Procedure1. Press PROG on the remote controller. The OutPut submenu appears.2. Press ENTER. The screen disp
MODEL 5081-T SECTION 7.0PROGRAMMING7.3 TEMP7.3.1 PurposeThis section describes how to do the following:1. Enable and disable automatic temperature co
ivMODEL 5081-T TABLE OF CONTENTSLIST OF FIGURES - CONT’DNumber Title Page5-1 Process Display Screen ...
54MODEL 5081-T SECTION 7.0PROGRAMMING1. Press PROG on the remote controller.2. Press NEXT until the diSPLAy submenu appears. Press ENTER.3. Press or
7.5 HARTIn multi-drop operation, polling addresses can be more conveniently set and debugging more conveniently performedusing the infrared remote con
567.6 SETUP CUSTThe Model 5081-T contains a curve fitting program that can create a second order curve for3 to 5 user supplied data points. If only tw
MODEL 5081-T SECTION 7.0PROGRAMMING7.7 RANGE7.7.1 PurposeThis section provides the steps to select automatic ranging or a specific fixed range of meas
58MODEL 5081-T SECTION 8.0FOUNDATION FIELDBUS OPERATIONSECTION 8.0FOUNDATION FIELDBUS OPERATIONThis section covers basic transmitter operation and sof
MODEL 5081-T SECTION 9.0 OPERATION WITH MODEL 375SECTION 9.0OPERATION WITH MODEL 3759.1 Note on Model 375 or 275 CommunicatorThe Model 375 or 275 Comm
609.3 Operation9.3.1 Off-line and On-line OperationThe Model 375 Communicator features off-line and on-line communications. On-line means the communic
MODEL 5081-T SECTION 9.0 OPERATION WITH MODEL 375--------------------------------------------------------------------------------5081-C/T 275, 375 Men
62MODEL 5081-T SECTION 9.0 OPERATION WITH MODEL 375Basic setup (continued)Device informationDistributorModelDev idTagDatePhysicl signl codeWrite prote
MODEL 5081-T SECTION 9.0 OPERATION WITH MODEL 375Diag override (continued)RTD fail [ON, OFF]Sense line open [ON, OFF]Temp high/low [ON, OFF]Signal con
1MODEL 5081-T SECTION 1.0DESCRIPTION AND SPECIFICATIONSSECTION 1.0DESCRIPTION AND SPECIFICATIONS• CHOICE OF COMMUNICATION PROTOCOL: HART or FOUNDATION
64MODEL 5081-T SECTION 9.0 OPERATION WITH MODEL 375Local DisplayAO LOI units [mA, %]Xmtr IDLoad Default Conf.ReviewPVPV AOPV LRVPV URV----------------
MODEL 5081-T SECTION 9.0 OPERATION WITH MODEL 3755081-C/T-FF/FI 375 Menu Tree-----------------------RESOURCEIdentificationMANUFACT_IDDEV_TYPEDEV_REVDD
66MODEL 5081-T SECTION 9.0 OPERATION WITH MODEL 375Transducer ErrorST_REVBLOCK_ERRAdditional transmitter statusFault history 0Fault history 1Fault his
MODEL 5081-T SECTION 9.0 OPERATION WITH MODEL 375ALERT_KEYHI_HI_LIMHI_HI_PRIHI_LIMHI_PRIIO_OPTSL_TYPELO_LO_LIMLO_LO_PRILO_LIMLO_PRIMODE_BLK: TargetMOD
68MODEL 5081-T SECTION 9.0 OPERATION WITH MODEL 375SIMULATE: Transducer StatusSIMULATE: Transducer ValueSIMULATE: Simulate En/DisableXD_SCALE: EU at 1
MODEL 5081-T SECTION 9.0 OPERATION WITH MODEL 375OUT_SCALE: EU at 0%OUT_SCALE: Units IndexOUT_SCALE: DecimalPV_SCALE: EU at 100%PV_SCALE: EU at 0%PV_S
70MODEL 5081-T SECTION 9.0 OPERATION WITH MODEL 375OnlineBK_CAL_IN: StatusBK_CAL_IN: ValueBK_CAL_OUT: StatusBK_CAL_OUT: ValueBLOCK_ERRBYPASSCAS_IN: St
MODEL 5081-T SECTION 9.0 OPERATION WITH MODEL 375DV_HI_ALM: Time StampDV_HI_ALM: SubcodeDV_HI_ALM: Float ValueDV_LO_ALM: UnacknowledgedDV_LO_ALM: Alar
72MODEL 5081-T SECTION 9.0 OPERATION WITH MODEL 375TRK_SCALE: Units IndexTRK_SCALE: DecimalTRK_IN_D: StatusTRK_IN_D: ValueTRK_VAL: StatusTRK_VAL: Valu
MODEL 5081-T SECTION 9.0 OPERATION WITH MODEL 375StdDevCap StdDevSchedulingDetailPhysical Device TagAddressDevice IDDevice RevisionAdvancedStack Capab
2MODEL 5081-T SECTION 1.0DESCRIPTION AND SPECIFICATIONS1.2 SPECIFICATIONS1.2.1 GENERAL SPECIFICATIONSEnclosure: Cast aluminum containing less than 6%
74MODEL 5081-T SECTION 9.0 OPERATION WITH MODEL 375Normal Rcv SAP LasTime Available SAP Src SinkNormal SAP Src SinkUrgent SAP Src SinkTime Available R
75MODEL 5081-T SECTION 10.0DIAGNOSIS AND TROUBLESHOOTINGSECTION 10.0DIAGNOSIS AND TROUBLESHOOTING10.1 OVERVIEWThe Model 5081-T transmitters automatica
76MODEL 5081-T SECTION 10.0DIAGNOSIS AND TROUBLESHOOTING-E< Absolute conductivity (µS/cm or mS/cm)"!DLS Sensor zero in air/1[[!/QPTU Sensor ce
MODEL 5081-T SECTION 10.0DIAGNOSIS AND TROUBLESHOOTING10.2 FAULT CONDITIONSThree classes of error conditions/problems are detectedand differentiated b
78MODEL 5081-T SECTION 10.0DIAGNOSIS AND TROUBLESHOOTING10.3 DIAGNOSTIC MESSAGESThe Model 5081-T transmitter’s diagnostics constantly monitor the con
79MODEL 5081-T SECTION 10.0DIAGNOSIS AND TROUBLESHOOTINGSYMPTOM ACTIONWrong temperature reading. Perform a temperature standardization. Verify sensor&
80MODEL 5081-T SECTION 10.0DIAGNOSIS AND TROUBLESHOOTING10.5 SYSTEMATIC TROUBLESHOOTINGIf the Quick Troubleshooting Guide does not resolve the error,
81MODEL 5081-T SECTION 10.0DIAGNOSIS AND TROUBLESHOOTING10.6 RTD RESISTANCE VALUESTable 10-4 is a ready reference of RTD resistance valuesat various t
82MODEL 5081-T SECTION 10.0DIAGNOSIS AND TROUBLESHOOTING10.7 WARNING AND FAULT MESSAGESThe Model 5081-T transmitter continuously monitors the sensor a
MODEL 5081-T SECTION 10.0DIAGNOSIS AND TROUBLESHOOTINGMessage Explanation See SectionOuEr rAnGE Over range, measurement exceeds display limit 10.8.1SE
MODEL 5081-T SECTION 1.0DESCRIPTION AND SPECIFICATIONS1.2.3 TRANSMITTER SPECIFICATIONS @ 25°CMeasured Range*: 50 to 2,000,000 µS/cm (see chart)Accura
84MODEL 5081-T SECTION 10.0DIAGNOSIS AND TROUBLESHOOTING10.8.3 CAL ErrorAt the end of the calibration step, the transmitter calculates the sensitivit
MODEL 5081-T SECTION 10.0DIAGNOSIS AND TROUBLESHOOTING10.8.7 OFFSEt ErrThe OFFSEt Err message appears if the zero offset (in mV) exceeds the programm
86MODEL 5081-T SECTION 11.0MAINTENANCESECTION 11.0MAINTENANCE11.3 TRANSMITTER MAINTENANCEPeriodically clean the transmitter window with house-hold amm
87MODEL 5081-T SECTION 11.0MAINTENANCEFIGURE 11-2. Exploded View of Model 5081-T TransmitterThree screws (part 13 in the drawing) hold the three circu
8812.1 OVERVIEWThis section is a general description of how the Model5081-T Transmitter operates. This section is for thoseusers who desire a greater
MODEL 5081-T SECTION 13.0RETURN OF MATERIALSECTION 13.0RETURN OF MATERIAL13.1 GENERAL.To expedite the repair and return of instruments, propercommunic
90This page intentionally left blank.
WARRANTYGoods and part(s) (excluding consumables) manufactured by Seller are warranted to be free from defects in workman-ship and material under norm