Emerson 3051S User Manual Page 49

  • Download
  • Add to my manuals
  • Print
  • Page
    / 84
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 48
49
Rosemount 3051
September 2014
www.rosemount.com
Functional specifications
Range and sensor limits
Table 7. 3051CD, 3051CG, 3051CF, and 3051L Range and Sensor Limits
Range
(1)
Minimum span Range and sensor limits
3051CD,
3051CG,
3051CF,
3051L
(2)
Upper
(URL)
Lower (LRL)
3051CD
differential
3051CF
Flowmeters
3051CG
gage
(3)
3051L
differential
3051L gage
(3)
0
0.10 inH
2
O
(0,24 mbar)
3.00 inH
2
O
(7,45 mbar)
-3.00 inH
2
O
(-7,45 mbar)
N/A N/A N/A
1
0.50 inH
2
O
(1,24 mbar)
25.00 inH
2
O
(62,16 mbar)
-25.00 inH
2
O
(-62,16 mbar)
-25.00 inH
2
O
(-62,16 mbar)
N/A N/A
2
1.67 inH
2
O
(4,15 mbar)
250.00 inH
2
O
(621,60 mbar)
-250.00 inH
2
O
(-621,60 mbar)
-250.00 inH
2
O
(-621,60 mbar)
-250.00 inH
2
O
(-621,60 mbar)
-250.00 inH
2
O
(-621,60 mbar)
3
6.67 inH
2
O
(16,58 mbar)
1000.00 inH
2
O
(2,48 bar)
-1000.00 inH
2
O
(-2,48 bar)
0.50 psia
(34,47 mbar)
-1000.00 inH
2
O
(-2,48 bar)
0.50 psia
(34,47 mbar)
4
2.00 psi
(137,89 mbar)
300.00 psi
(20,68 bar)
-300.00 psi
(-20,68 bar)
0.50 psia
(34,47 mbar)
-300.00 psi
(-20,68 bar)
0.50 psia
(34,47 mbar)
5
13.33 psi
(919,01 mbar)
2000.00 psi
(137,89 bar)
- 2000.00 psi
(-137,89 bar)
0.50 psia
(34,47 mbar)
N/A N/A
(1) Range 0 only available with 3051CD. Range 1 only available with 3051CD, 3051CG, or 3051CF. inH
2
O referenced at 68 degrees Fahrenheit.
(2) For outputs options W and M, minimum span are: range 1 - 0.50 inH
2
O (1,24 mbar), range 2 - 2.50 inH
2
O (6,21 mbar), range 3 - 10.00 inH
2
O
(24,86 mbar), range 4 - 3.00 psi (0,21 bar), range 5 - 20.00 psi (1,38 bar).
(3) Assumes atmospheric pressure of 14.7 psig.
Table 8. 3051CA and 3051T Range and Sensor Limits
Range
3051CA
Range
3051T
Minimum span
(1)
Range and sensor limits Minimum span
(1)
Range and sensor limits
Lower
(2)
(LRL) (gage)
Upper
(URL)
Lower
(LRL)
Upper
(URL)
Lower
(LRL)
(absolute)
1
0.30 psi
(20,68 mbar)
30 psia
(2,06 bar)
0 psia
(0 bar)
1
0.30 psi
(20,68 mbar)
30.00 psi
(2,06 bar)
0 psia
(0 bar)
-14.70 psig
(-1,01 bar)
2
1.00 psi
(68,94 mbar)
150 psia
(10,34 bar)
0 psia
(0 bar)
2
1.00 psi
(68,94 mbar)
150.00 psi
(10,34 bar)
0 psia
(0 bar)
-14.70 psig
(-1,01 bar)
3
5.33 psi
(367,49 mbar)
800 psia
(55,15 bar)
0 psia
(0 bar)
3
5.33 psi
(367,49 mbar)
800.00 psi
(55,15 bar)
0 psia
(0 bar)
-14.70 psig
(-1,01 bar)
4
26.67 psi
(1,83 bar)
4000 psia
(275,79 bar)
0 psia
(0 bar)
4
26.67 psi
(1,83 bar)
4000.00 psi
(275,79 bar)
0 psia
(0 bar)
-14.70 psig
(-1,01 bar)
5 N/A N/A N/A 5
2000.00 psi
(137,89 bar)
10000.00 psi
(689,47 bar)
0 psia
(0 bar)
-14.70 psig
(-1,01 bar)
(1) For output options W and M, minimum span are: range 2 - 1.50 psi(0,10 bar), range 3 - 8.00 psi (0,55 bar), range 4 - 40.00 psi (2,75 bar), range 5 for 3051T
- 2000.00 psi (137,89 bar)
(2) Assumes atmospheric pressure of 14.7 psig.
Page view 48
1 2 ... 44 45 46 47 48 49 50 51 52 53 54 ... 83 84

Comments to this Manuals