www.Fisher.comFisherr 4200 Electronic Position TransmittersThe instrument can sense the position of rotary orsliding‐stem valves, vents, dampers or ot
4200 TransmittersD200357X012Product Bulletin62.3:4200July 2014 10Table 7. Hazardous Area Classifications—ATEXCertificate Type Certification Obtained E
4200 TransmittersD200357X012Product Bulletin62.3:4200July 2014 11
4200 TransmittersD200357X012Product Bulletin62.3:4200July 2014 12Emerson Process Management Marshalltown, Iowa 50158 USASorocaba, 18087 BrazilChatham,
4200 TransmittersD200357X012Product Bulletin62.3:4200July 2014 2SpecificationsAvailable ConfigurationsSee table 1Input Signal SourceJ Standard single
4200 TransmittersD200357X012Product Bulletin62.3:4200July 2014 3Specifications (continued)Electrical ClassificationCSA—Intrinsically Safe, Explosion-p
4200 TransmittersD200357X012Product Bulletin62.3:4200July 2014 4Table 3. Power Supply Requirements and Wiring ConnectionsTRANSMITTER TERMINALVOLTAGE (
4200 TransmittersD200357X012Product Bulletin62.3:4200July 2014 5Figure 2. Fisher 4210 Transmitter DetailsW9275W4675‐1TERMINAL BLOCKNUMBER 1TRANSMITTER
4200 TransmittersD200357X012Product Bulletin62.3:4200July 2014 6Figure 3. DimensionsPORT1/2 NPT(2 PLACES)TRANSMITTERCOMPARTMENTFIELD WIRINGCOMPARTMENT
4200 TransmittersD200357X012Product Bulletin62.3:4200July 2014 7Long‐Stroke Position TransmitterLong‐Stroke Sliding‐Stem Valve—The transmitter ismount
4200 TransmittersD200357X012Product Bulletin62.3:4200July 2014 8Figure 4. Mounting Dimensions51(2.00)48(1.88)108(4.25)271(10.69)204(8.00)86(3.40)10(0.
4200 TransmittersD200357X012Product Bulletin62.3:4200July 2014 9Table 4. Fisher 4211 and 4221 Electronic Position Transmitter EMC Summary Results—Immu