Product Data SheetPDS 103-102.A01June, 2004MicroCEM™Visit our website at www.processanalytic.comOn-line ordering available.Continuous Emissions Monito
Page 10 Model Description USHS MicroCEM Probe/Sample Handling System (USHS) Level 1 Standard Interior Sample Handling System, In Fiberglass Enclo
Page 11
Emerson Process ManagementRosemount Analytical Inc.Process Analytic Division1201 North Main StreetOrrville, OH 44667-0901 USAT 330.682.9010Toll Fre
Page 2HousingThe housing is equipped with a weather-resistantNEMA 4X-rated enclosure and can handle ambienttemperatures between -30 to 50°C. The hous
Page 3Analysis EnclosureThe MicroCEM design includes an analysis enclosurewhich can be located at the sampling location at thebottom of the stack or i
• The MicroCEM is equipped with a manifold with three,two-way closed solenoid valves that will direct thecalibration gases into the system. The three
Page 5Power: Universal power supply 85 to 125VAC, 50 to 60 Hz, +/-10%1000 watts maximum at start up,500 watts nominalMicroprocessor: Intel Celeron pro
Page 6Power: Universal power supply 85 to125 VAC, 50 to 60 Hz, +/-10%750 watts maximum atstart up. 500 watts nominalMounting: Customer flange mount(2
Page 7EXAMPLE MENU SCREENSSettings 2LimitsFile Tools AdvancedAuto Cal. Freq.Auto Cal.RangeO2 Output RangeRange 225.00 %Range 15.00
Page 8 Model Description UCEMS MicroCEM PC104-Controlled O2/CO/NOx CEM in a NEMA 4X Enclosure (UCEMS) Level 1 Channel 1 11 Single Channel Parama
Page 9 Model Description UCEMTS MicroCEM PC104-Controlled O2/CO/NOx Time-Sharing CEM in a NEMA 4X Enclosure (UCEMTS) Level 1 Application Configur